We report a colorimetric method for determining lithium in blood serum without sample pretreatment or solvent-extraction steps. The method is based on a novel chromogenic ionophore that exhibits exceptionally high selectivity (much greater than 4000:1) for lithium over sodium. The standard curve for the method is linear up to 3.5 mmol/L and exceeds the therapeutic range for lithium. The results for 57 patients' samples correlated well with results obtained with an Instrumentation Laboratory flame photometer (r = 0.97).
We report a new single-reagent colorimetric assay for the determination of calcium in serum. The assay is based on a novel chromogenic tetracarboxylic acid chelating agent. The method has several advantages over the o-cresolphthalein complexone (CPC) method, including improved linearity, better accuracy, increased stability, and lower working pH (9.0). The calibration curve for the assay is linear up to 4 mmol/L. The results for 47 patients' samples correlated well with results obtained with an Instrumentation Laboratory atomic absorption spectrophotometer. The mean (+/- SD) calcium values for the new method were 2.27 +/- 0.06 mmol/L and 2.26 +/- 0.05 mmol/L for the comparison method. The on-system stability of the reagents is 1 week vs 8 h for the CPC method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.