Background
Tight control of tidal volume using accurate monitoring may improve neonatal outcomes. However, respiratory function monitors incorporated in current anesthetic workstations are generally inaccurate at tidal volumes used for infants.
Aims
To determine if a specific respiratory function monitor for neonatal infants improved expired tidal volume delivery during anesthesia.
Method
Infants <3 months old requiring intubation for surgery in the operating theater were studied. After intubation a Phillips NM3, Acutronic Florian, or Novametrix Ventcheck Respiratory Function Monitor was integrated into the circuit, and clinicians given access to the display for the duration of anesthesia. Breath‐to‐breath expired tidal volume delivery, leak, and delivered pressure were recorded, with cardiorespiratory parameters. These were compared with a matched control group with clinicians blinded to respiratory function monitor display.
Results
A total of 10 055 and 2569 inflations were measured in the respiratory function monitor visible (n = 32) and masked (n = 33) groups, respectively, with mean (standard deviation) delivered expired tidal volume 7.5 (2.4) mL/kg and 7.7 (3.0) mL/kg, respectively; mean difference (95% confidence interval) −0.2 (−1.1, 0.8) mL/kg (Welch's t test). In the visible group, 55.6% of expired tidal volumes were between 4 and 8 mL/kg compared to 51.7% in the masked group; relative benefit (95% confidence interval), 1.08 (1.03, 1.12). Expired tidal volume was less likely to be <4 mL/kg in the visible group compared to masked group; 6.4% vs 9.8%, 1.53 (1.33, 1.76). The use of a respiratory function monitor also reduced the number of inflations >10 mL/kg; 13.0% vs 22.0%, 1.11 (1.09, 1.14).
Conclusion
Tidal volumes <4 mL/kg and >10 mL/kg are frequently delivered during neonatal anesthesia. The inclusion of an accurate respiratory function monitor may reduce the risk of exposure to potentially harmful tidal volumes.
In the original article's XML, we stated co-author Dr Rachael C. Heath Jeffery's name incorrectly (we entered "Heath" as part of her given name rather than her family name). We've corrected the XML (which automatically fixed the previously incorrect Pubmed indexing).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.