Extended-range BSS, TEPCs, and the WENDI-II enable accurate measurements of stray neutrons while other rem-counters are not appropriate considering the high-energy range of neutrons involved in proton therapy.
The purpose of this work was an evaluation of organ doses and effective doses from three verification techniques in Image-Guided Radiotherapy: from kilovoltage (kV) cone beam computed tomography (CBCT) scans, from two orthogonal kV images and from two orthogonal megavoltage (MV) images for two different treatment sites: pelvis and head and neck (H&N). For comparison reasons, organ doses and effective doses from prostate and H&N radiotherapy were also evaluated. Measurements of organ doses were performed in a male anthropomorphic Rando phantom by means of thermoluminescent dosemeters. In this investigation, measured organ doses from one CBCT scan, from two MV images and from two kV images of pelvis represent typically 1-6, 1-10 and 0.05-1 %, respectively, of organ doses resulting from one fraction of prostate radiotherapy. The maximum effective doses from CBCT scans, kV images and MV images of pelvis are 5.6, 0.8 and 11.9 mSv, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.