BackgroundFor electroporation-based therapies, accurate modeling of the electric field distribution within the target tissue is important for predicting the treatment volume. In response to conventional, unipolar pulses, the electrical impedance of a tissue varies as a function of the local electric field, leading to a redistribution of the field. These dynamic impedance changes, which depend on the tissue type and the applied electric field, need to be quantified a priori, making mathematical modeling complicated. Here, it is shown that the impedance changes during high-frequency, bipolar electroporation therapy are reduced, and the electric field distribution can be approximated using the analytical solution to Laplace's equation that is valid for a homogeneous medium of constant conductivity.MethodsTwo methods were used to examine the agreement between the analytical solution to Laplace's equation and the electric fields generated by 100 µs unipolar pulses and bursts of 1 µs bipolar pulses. First, pulses were applied to potato tuber tissue while an infrared camera was used to monitor the temperature distribution in real-time as a corollary to the electric field distribution. The analytical solution was overlaid on the thermal images for a qualitative assessment of the electric fields. Second, potato ablations were performed and the lesion size was measured along the x- and y-axes. These values were compared to the analytical solution to quantify its ability to predict treatment outcomes. To analyze the dynamic impedance changes due to electroporation at different frequencies, electrical impedance measurements (1 Hz to 1 MHz) were made before and after the treatment of potato tissue.ResultsFor high-frequency bipolar burst treatment, the thermal images closely mirrored the constant electric field contours. The potato tissue lesions differed from the analytical solution by 39.7 ± 1.3 % (x-axis) and 6.87 ± 6.26 % (y-axis) for conventional unipolar pulses, and 15.46 ± 1.37 % (x-axis) and 3.63 ± 5.9 % (y-axis) for high- frequency bipolar pulses.ConclusionsThe electric field distributions due to high-frequency, bipolar electroporation pulses can be closely approximated with the homogeneous analytical solution. This paves way for modeling fields without prior characterization of non-linear tissue properties, and thereby simplifying electroporation procedures.
Irreversible electroporation (IRE) describes a cellular response to electric field exposure, resulting in the formation of nanoscale defects that can lead to cell death. While this behavior occurs independently of thermally-induced processes, therapeutic ablation of targeted tissues with IRE uses a series of brief electric pulses, whose parameters result in secondary Joule heating of the tissue. Where contemporary clinical pulse protocols use aggressive energy regimes, additional evidence is supplementing original studies that assert care must be taken in clinical ablation protocols to ensure the cumulative thermal effects do not induce damage that will alter outcomes for therapies using the IRE non-thermal cell death process for tissue ablation. In this letter, we seek to clarify the nomenclature regarding IRE as a non-thermal ablation technique, as well as identify existing literature that uses experimental, clinical, and numerical results to discretely address and evaluate the thermal considerations relevant when applying IRE in clinical scenarios, including several approaches for reducing these effects. Existing evidence in the literature describes cell response to electric fields, suggesting cell death from IRE is a unique process, independent from traditional thermal damage. Numerical simulations, as well as preclinical and clinical findings demonstrate the ability to deliver therapeutic IRE ablation without occurrence of morbidity associated with thermal therapies. Clinical IRE therapy generates thermal effects, which may moderate the non-thermal aspects of IRE ablation. Appropriate protocol development, utilization, and pulse delivery devices may be implemented to restrain these effects and maintain IRE as the vastly predominant tissue death modality, reducing therapy-mitigating thermal damage. Clinical applications of IRE should consider thermal effects and employ protocols to ensure safe and effective therapy delivery.
The internally-cooled bipolar applicator offers advantages that could improve clinical outcomes. Thermally mitigating internal perfusion technology reduced tissue temperatures and electric current while maintaining similar lesion sizes.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.