The radiation detection properties of semiconductor detectors made of 4H silicon carbide were evaluated. Both Schottky and p-n junction devices were tested. Exposure to alpha particles from a 238Pu source led to robust signals from the detectors. The resolution of the Schottky S i c detector was 5.8% (FWHM) at an energy of 294 keV, while that of the p-n junction was 6.6% (EWHM) at 260 keV.No effect of temperature in the range of 22 to 89 "C was observed on the characteristics of the Pu alpha-induced signal from the S i c detector. In addition, testing in a gamma field of 10,000 rad-Si h" showed that the alpha-induced signal was separable from the gamma signal. 238
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.