The preoperative pure-tone threshold in the implanted ear was between 80 and 115 dB in the frequency range of 500 to 2000 Hz. On the contralateral side the mean threshold was 10 dB better. The mean score for the number test was 11.3%, for the monosyllables below 5%. Postoperatively speech understanding improved significantly up to 97% for numbers and 48% for monosyllables (tested with the Freiburger Speech Test). Cochlear implantation also benefits patients with residual hearing and some speech understanding with hearing aids under optimum conditions. The speech understanding scores must be below certain limits. General selection criteria cannot yet be specified. The individual decision must be based upon several criteria, especially the speech understanding scores in quiet and noise under optimum conditions. A prospective study is needed to develop generally applicable criteria.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.