The AAPM Task Group 21 protocol is the latest amongst the protocols introduced in recent years covering the dosimetry of high-energy x-ray and electron beams used in radiotherapy. We have found that the C lambda value for 15-MV x rays requires 1.6% upward adjustment, whereas the CE values for electrons in the range 7-18 MeV require downward adjustment of 1%-3% compared to our previous values. The adjustment values given in this paper pertain to the specific measuring systems and high-energy beams considered here and would not necessarily be applicable to other systems. The dose to medium, calculated for different modalities, was verified by thermoluminescent dosimetry and the results were found to be consistent. These values of C lambda and CE have also been compared with those derived using other protocols, particularly the Nordic Association of Clinical Physics 80 and 81, and the discrepancies discussed.
Using a focal spot of 40-120 microns, the new mammographic system provides markedly improved direct radiographic magnification. With reduced radiation exposure at times, there is a significant gain in diagnostic information compared with current standard systems.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.