A reversed-phase high-performance liquid chromatographic method (HPLC) with diode-array detection is developed and validated for the quantitative determination of formaldehyde in a drug substance. Formaldehyde (HCHO) is reacted with 2,4-dinitrophenylhydrazine (DNPH) to form a Schiff base (HCHO-DNPH derivatization product), which has an absorbing maximum (lambda max) at 360 nm. The HPLC method employs a C8, 3-microm particle size analytical column (150 mm x 4.6 mm), 15-microL injection volume, column temperature controlled at 30 degrees C, detection at 360 nm, and a water-acetonitrile (55:45, v/v) mobile phase at a flow rate of 1 mL/min. These conditions resolve the HCHO-DNPH product from unreacted DNPH, the drug substance and related impurities, as well as diluent peaks within 20 min. The retention time of the HCHO-DNPH product is approximately 6.4 min. The method is linear, accurate in the specified range (0.33-333 ppm), and robust based on analyte (HCHO-DNPH derivatization product) stability in standard and sample. Detection limit is 0.03 ng (0.1 ppm).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.