SUMMARY
The scan speed limit of atomic force microscopes has been calculated. It is determined by the spring constant of the cantilever k, its effective mass m, the damping constant D of the cantilever in the surrounding medium and the stiffness of the sample. Techniques to measure k, k/m and D/m are described. In liquids the damping constant and the effective mass of the cantilever increase. A consequence of this is that the transfer function always depends on the scan speed when imaging in liquids. The practical scan speed limit for atomic resolution in vacuum is 0·1 μm/s while in water it increases to about 2 μm/s due to the additional damping of cantilever movements. Sample stiffness or damping of cantilever movements by the sample increase these limits. For soft biological materials imaged in water at a desired resolution of 1 nm the scan speed should not exceed 2 μm/s.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.