The evaluation of oxomemazine hydrochloride (OXCl) prepared in carbon paste electrodes is reported in detail. There are many practical techniques and experiments for electrode preparation one of them is mixing of the ion-pair with ion pairing agent. Prepared electrodes gave satisfactory data in ranging from (0.018 to 18.35) mg based on Nernstian cationic slopes, the used electrodes refer to the standard data obtained at (slope 58.7±2.1 mV decade-1) by using sodium tetraphenylborate (NaTPB) and rapid response time equal to 15 s and lifetime approximately equal to (4 weeks). The prepared electrodes can be counted as end point indicator electrode for potentiometric titration of oxomemazine Hydrochloride OXCl resulting in a high percentage of accuracy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.