In this paper, a versatile, re-programmable, current-input bioinstrumentation board is presented for electrochemical amperometric measurements. The proposed instrument has been fabricated on a six layer printed circuit board (PCB) and exploits dual switched-capacitor (SC) integration and sampleand-hold (SH) techniques. It comprises off-the-shelf switch and amplifier ICs and a commercially available FPGA-based DSP unit for digital signal control and synchronisation. It features eight amperometric channels, has a dynamic current range of 100dB, can be powered-up by a USB port or a 5V battery and is portable, with dimensions of 110×110 mm 2. An onboard digital-to-analog converter (DAC) combined with standard three-electrode potentiostats can provide precise, programmable biasing voltages to eight amperometric biosensors simultaneously. Validation of the robustness and accuracy of the proposed system is demonstrated by proof-of-concept amperometric measurements using a high-precision Keithley 6221 current source and NaCl solution on a PCB-based sensor.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.