Sulfamethazine (SMZ) and trimethoprim (TMP) are antibacterials used in veterinary practice. This paper describes a method for their determination in veterinary liquid feed premixes that is based on liquid chromatography with diode array detection. Gradient elution with methanol and ammonium acetate achieved excellent separation of the two analytes within 15 min without any interference from the matrix. Absorbance of the column effluent was monitored at 264 nm for SMZ and at 230 nm for TMP. Detailed analyses of the uncertainties of determinations afford estimated expanded uncertainties of, respectively, 0.2 and 0.1 w/v % for typical SMZ and TMP concentrations of 10.7 and 2.1 w/v %, respectively. At the lower end of the calibrated range of the method, the dominant source of uncertainty is the preparation of standards and the construction of the calibration line.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.