Choline was determined in infant formula by ion chromatography with suppressed conductivity detection. Samples were digested with 1M hydrochloric acid, filtered, diluted, and injected into the chromatographic system. Choline and the alkali and alkaline earth metals were separated on a high-resolution cation-exchange column and detected by suppressed conductivity. The method was linear between 2 and 200 mg/L (r2 = 0.9999), the concentration range of the diluted samples. This method accurately determined choline in powdered, concentrated, and ready-to-feed infant formulas. Recoveries of choline spikes into powdered infant formula at approximately 1, 0.8, 0.5, and 0.2 times the labeled value ranged from 85 to 114%. This method had good agreement for 8 blind duplicates. The values determined for these samples, which were used in an AOAC collaborative study of an enzymatic method, were consistent with the values determined by the enzymatic method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.