Fibrinogen is one of the first proteins to be depleted in heavily bleeding patients. In this study, we have developed a new paper-based diagnostic to quantify the fibrinogen concentration in blood at room temperature.
The ability to diagnose and treat
critically bleeding patients
can save more than 2 million lives a year. Diagnosing hypofibrinogenemia
is essential in these patients. Recently, with the development of
new handheld diagnostics, fibrinogen concentration can be measured
rapidly at the point of care. However, these diagnostics can only
work with plasma and hence need blood cells to be separated before
use. In this study, we demonstrate a handheld fibrinogen diagnostic
that works with whole blood. The test works by (1) forming a premixed
droplet of a whole blood sample and thrombin solution on a solid surface,
(2) allowing it to clot, and (3) dropping a paper strip on top. The
further that blood moves down the strip, the lower the fibrinogen
concentration. The diagnostic can easily measure plasma fibrinogen
concentrations below 1.6 g/L for blood samples with hematocrits between
40 and 50%. Furthermore, diluting blood samples not only increases
the test’s sensitivity but also eliminates the effect of hematocrit
and thrombin inhibitors. The test can be completed in 3–4 min,
making it suitable for diagnosing early hypofibrinogenemia and allowing
for fibrinogen replacement therapy in critically bleeding patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.