Heparin is widely used as an anticoagulant for the treatment and prevention of various thrombotic diseases. However, due to its high anionic charge, heterogeneity in size distribution and high polarity, its analysis is very challenging. In this paper, a novel method based on size-exclusion chromatography (SEC) for quantitative determination of intact heparin in pharmaceuticals is presented. Analyses were performed on a BioSep-SEC-S 2000 column with Larginine solution at pH 6.5 as mobile phase and UV detection at 210 nm. The proposed method was found to be selective, linear (R 2 > 0.997) over the concentration range of 3.1 to 1222 µg mL -1 , with a limit of detection of 1.0 µg mL -1 . Intraday and inter-day precision were below 5.1 % and inaccuracy expressed as bias did not exceed 6.5 %. The reported method is simple, selective, sensitive, and requires no laborious sample preparation, which makes it appropriate for routine quantitative analysis of heparin in pharmaceuticals.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.