Impregnated silica TLC plates with L-(-)-serine and L-(-)-threonine and a mixture of L-(-)-serine and L-(-)-threonine (1:1) as chiral selectors were prepared to use as chiral stationary phases (CSPs) in thin layer chromatography. The resolution of the enantiomers of 2-arylpropionic drugs, including ibuprofen, ibuproxam, ketoprofen, pranoprofen, benoxaprofen, flurbiprofen and tiaprofenic acid was investigated on these CSPs. A mobile phase system of acetonitrile-methanol-water (16:4:0.5, v/v/v) was used. The spots were detected with iodine vapours and the detection limits were found to range between 0.25 and 0.5 micro g/mL for all racemic compounds investigated. The effect of temperature, pH and concentration of the impregnating chiral selectors on resolution has been studied.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.