For the rapid and simple diagnosis of schistosomiasis, a reagent strip assay for detection of circulating cathodic antigen (CCA) in urine was developed. The test was based on a previously developed sandwich enzyme-linked immunosorbent assay (ELISA) using a combination of two anti-CCA monoclonal antibodies. For the reagent strip assay, as the capture matrix, monoclonal antibody was coated onto a nitrocellulose membrane and mounted on polyvinyl chloride strips. Urine samples were then tested in an assay consisting of a combined incubation step of the urine sample and biotinylated detecting antibody followed by incubation in streptavidin peroxidase and a subsequent staining. The specificity and the sensitivity of the assay, as determined with urine samples of 61 uninfected controls and 67 Schistosoma mansoni-infected individuals, were 96.7 and 95.5%, respectively. The results of the reagent strip assay compared very well with microscopical parasitological diagnosis by the standard Kato-Katz method for the same individuals. The reagent strip test has a lower detection limit of 1 ng of CCA per ml and can be completed in 75 min. By the inclusion of two reference bands on the strips, standardized reading could be achieved. This reagent strip assay is a promising tool for qualitative diagnosis of S. mansoni infections in control programs.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.