Additive manufacturing is a very promising technique for the production of complex multipinhole collimators and may also be used for producing other complex collimators. The cost of this technique is only related to the amount of powder needed and the time it takes to have the collimator built. The timeframe from design to collimator production is significantly reduced.
In single-photon emission computed tomography (SPECT), multi-pinhole collimation is often employed nowadays. Most multi-pinhole collimators avoid overlap (multiplexing) of the projections on the detector. This can be done by using additional shielding or by spacing the pinholes far enough apart. Using additional shielding has the drawback that it increases weight, design complexity and cost. Spacing the pinholes far enough apart results in sub-optimal detector usage, the valuable detector area is not entirely used. This is due to the circular projections of pinholes on the detector; these ellipses can not be tiled with high detector coverage. To overcome this we designed a new pinhole geometry, the lofthole, that has a rectangular projection on the detector. The lofthole has a circular aperture and a rectangular entrance/exit opening. Sensitivity formulae have been derived for pinholes and loftholes. These formulae take the penumbra effect into account; the proposed formulae do not take penetration into account. The derived formulae are valid for geometries where the field-of-view and the sensitivity of the aperture are solely limited by the exit window. A flood map measurement was performed to compare the rectangular projection of a lofthole with the circular projection of a pinhole. Finally, measurements were done to compare the amount of penetration of pinholes with the amount of penetration of a lofthole. A square lofthole collimator has less penetration than a knife-edge pinhole collimator that irradiates the same rectangular detector area with full coverage. A multi-lofthole collimator allows high detector coverage without using additional shielding. An additional advantage is the lower amount of penetration.
Currently, clinical brain single photon emission computed tomography (SPECT) is mostly performed using rotating dual-head gamma cameras equipped with low-energy-high-resolution parallel-beam collimators (LEHR PAR). The resolution of these systems is rather poor (8-10 mm) and the rotation of the heavy gamma cameras can introduce misalignment errors. Therefore, we designed a static full-ring multi-lofthole brain SPECT insert for an existing ring of LaBr3 (5% Ce) detectors. The novelty of the design is found in the shutter mechanism that makes the system very flexible and eliminates the need for rotating parts. A stationary SPECT insert is not only more robust, it is also easier to integrate in a magnetic resonance imaging system (MRI) for simultaneous SPECT-MRI. The target spatial resolution of our design is 6 mm. In this study we used analytical calculations to optimize the collimator for an existing ring of LaBr3 (5% Ce) detectors. We fixed the target spatial resolution at 6 mm in the center of the field-of-view and maximized the volume sensitivity by changing the collimator radius, the aperture and the number of loftholes. Based on these optimal parameters we simulated phantom data and evaluated the image quality of our multi-lofthole system. We simulated a noiseless uniform and Defrise phantom to assess artifacts and sampling completeness and a noiseless hot-rod phantom to assess the reconstructed spatial resolution. We visually evaluated a simulated noisy Hoffman phantom with two lesions. Then, we evaluated the non-prewhitening matched filter signal-to-noise ratio (NPW-SNR) in two lesion detectability phantoms: one with hot lesions and one with cold lesions. Finally, a contrast-to-noise (CNR) study was performed on a phantom with both hot and cold lesions of different sizes (6-16 mm). All results were compared to a LEHR PAR system. The optimization resulted in a final collimator design with a volume sensitivity of 1.55 × 10(-4) cps Bq(-1), which is 2.5 times lower than the sensitivity of a dual-head system with LEHR PAR collimators. Spatial resolution, on the other hand, has clearly improved compared to LEHR PAR: with the multi-lofthole system we successfully reconstructed 4 mm hot rods. Although this improved resolution did not result in an unambiguous improvement in CNR or NPW-SNR, we believe that the flexibility of the shutter mechanism opens interesting perspectives toward time-multiplexing and integration with MRI.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.