Hydroxyapatite cement (BoneSource®) and brushite calcium phosphate cement (chronOS™ Inject) were tested for fixation of glass ceramic implants (Bioverit®) in experimentally created cranial defects in 24 adult New Zealand White rabbits. Aim of the in vivo study was to assess and compare the biocompatibility and osseointegration of the implanted materials. Macroscopic and histological evaluations were performed 1 month, 3 months, and 6 months postoperatively. All implanted materials were well tolerated by the surrounding tissue. Both bone cements exhibited osteoconductive properties. Differences could be detected regarding to the rates of cement resorption and new bone formation. The brushite cement was resorbed faster than the hydroxyapatite cement. The chronOS™ Inject samples exhibited a higher rate of connective tissue formation and an insufficient osseointegration. BoneSource® was replaced by bone with minimal invasion of connective tissue. New bone formation occurred faster compared to the chronOS™ Inject group. Bioverit® implants fixed with BoneSource® were successfully osseointegrated.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.