Purpose
To present a theoretical basis for noninvasively characterizing in vivo fluid-mechanical energy losses, and to apply it in a pilot study of patients known to express abnormal aortic flow patterns.
Methods
4D flow MRI was used to characterize laminar viscous energy losses in the aorta of normal controls (n=12, age=37±10), patients with aortic dilation (n=16, age=52±8), and patients with aortic valve stenosis matched for age and aortic size (n=14, age=46±15), using a relationship between the 3D velocity field and viscous energy dissipation.
Results
Viscous energy loss was significantly elevated in the thoracic aorta for patients with dilated aorta (3.6±1.3 mW, p=0.024) and patients with aortic stenosis (14.3±8.2 mW, p<0.001) compared to healthy volunteers (2.3±0.9 mW). The same pattern of significant differences were seen in the ascending aorta, where viscous energy losses in patients with dilated aortas (2.2±1.1 mW, p=0.021) and patients with aortic stenosis (10.9±6.8 mW, p<0.001) were elevated compared to healthy volunteers (1.2±0.6 mW).
Conclusion
This technique provides a capability to quantify the contribution of abnormal laminar blood flow to increased ventricular afterload. In this pilot study, viscous energy loss in patient cohorts was significantly elevated and indicates that cardiac afterload is increased due to abnormal flow.
Purpose
To improve velocity–to-noise ratio (VNR) and dynamic velocity range at 4D flow MRI by using dual-velocity encoding (dual-venc) with k-t GRAPPA acceleration.
Materials and Methods
A dual-venc 4D flow MRI sequence with k-t GRAPPA acceleration was developed using a shared reference scan followed by three-directional low- and high-venc scans (TR/TE/FA=6.1ms/3.4ms/15°, temporal/spatial resolution=43.0ms/1.2×1.2×1.2mm3). The high-venc data was used to correct for aliasing in the low-venc data, resulting in a single dataset with the favorable VNR of the low-venc but without velocity aliasing. The sequence was validated at a 3 Tesla MRI Scanner in phantom experiments and applied in 16 volunteers to investigate its feasibility for assessing intracranial hemodynamics (net flow and peak velocity) at the major intracranial vessels. In addition, image quality and image noise was assessed in the in-vivo acquisitions.
Results
All 4D flow MRI scans were acquired successfully with an acquisition time of 20±4min. The shared reference scan reduced the total acquisition time by 12.5% compared to two separate scans. Phantom experiments showed 51.4% reduced noise for dual-venc compared to high-venc and an excellent agreement of velocities (ρ=0.8, p<0.001). The volunteer data showed decreased noise in dual-venc data (54.6% lower) compared to high-venc, and improved image quality, as graded by two observers: less artifacts (P<0.0001), improved vessel conspicuity (P<0.0001), and reduced noise (P<0.0001).
Conclusion
Dual-venc 4D flow MRI exhibits the superior VNR of the low-venc acquisition and reliably incorporates low- and high-velocity fields simultaneously. In-vitro and in-vivo data demonstrate improved flow visualization, image quality and image noise.
Background: The effective orifice area (EOA) estimated by transthoracic Doppler echocardiography (TTE) via the continuity equation is commonly used to determine the severity of aortic stenosis (AS). However, there are often discrepancies between TTE-derived EOA and invasive indices of stenosis, thus raising uncertainty about actual definite severity. Cardiovascular magnetic resonance (CMR) has emerged as an alternative method for non-invasive estimation of valve EOA. The objective of this study was to assess the concordance between TTE and CMR for the estimation of valve EOA. Methods and results: 31 patients with mild to severe AS (EOA range: 0.72 to 1.73 cm
Objectives
To comprehensively and quantitatively analyze flow and vorticity in the right heart of patients after repair of tetralogy of Fallot (rTOF) compared with healthy volunteers.
Methods
Time-resolved flow-sensitive 4D MRI was acquired in 24 rTOF patients and 12 volunteers. Qualitative flow evaluation was based on consensus reading of two observers. Quantitative analysis included segmentation of the right atrium (RA) and ventricle (RV) in a 4-chamber view to extract volumes and regional hemodynamic information for computation of regional mean and peak vorticity.
Results
Right heart intra-atrial, intraventricular and outflow tract flow patterns differed considerably between rTOF patients and volunteers. Peak RA and mean RV vorticity was significantly higher in patients (p=0.02/0.05). Significant negative correlations were found between patients’ maximum and mean RV and RA vorticity and ventricular volumes (p<0.05). The main pulmonary artery (MPA) regurgitant flow was associated with higher RA and RV vorticity, which was significant for RA maximum and RV mean vorticity (p=0.01/0.03).
Conclusion
The calculation of vorticity based on 4D flow data is an alternative approach to assess intracardiac flow changes in rTOF patients compared with qualitative flow visualization. Alterations in intracardiac vorticity could be relevant with regard to the development of RV dilation and impaired function.
Purpose
To investigate age-related changes in peak systolic aortic 3D velocity and wall shear stress (WSS) in healthy controls and to investigate the importance of age-matching for 3D mapping of abnormal aortic hemodynamics in bicuspid aortic valve disease (BAV).
Methods
4D flow MRI (fields strengths=1.5 – 3T; resolution=2.2–3.9x1.7–2.6x2.2–4.0mm3; venc=150–250cm/s; TE/TR/FA=2.3–2.8ms/4.7–5.4ms/7–15°) was performed in 56 controls (age range: 19–78 years) and in two BAV patient groups each consisting of 10 subjects (group 1: 20–29 years, group 2: 52–57 years). Heat maps showing abnormal 3D velocity and WSS were created for the BAV patients by comparison with an age-matched and with an unmatched control group. The fraction of the aorta exposed to abnormal velocity/WSS was calculated relative to the total aortic volume/surface.
Results
Significant inverse relationships between age and healthy velocity/WSS were found (R2=0.32/0.39, P<0.001). For BAV group 1, abnormally elevated velocity/WSS was overestimated when compared with older controls (51–60 years) than when correctly age-matched (~25±14% vs. ~8±5%). For BAV group 2, abnormally decreased velocity/WSS was overestimated when compared with younger controls (21–30 years) than when correctly age-matched (~9±7% vs. 1±1%).
Conclusion
Significant correlations exist between age and peak systolic velocity and WSS. Therefore, robust age-matching is important when creating abnormal 3D aortic velocity and WSS maps for patients with BAV.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.