The purpose of this study was to assess the patient exposure and staff eye dose levels during implantation procedures for all types of pacemaker therapy devices performed under fluoroscopic guidance and to investigate potential correlation between patients and staff dose levels. The mean eye dose during pacemaker/defibrillator implementation was 12 µSv for the first operator, 8.7 µSv for the second operator/nurse and 0.50 µSv for radiographer. Corresponding values for cardiac resynchronisation therapy procedures were 30, 26 and 2.0 µSv, respectively. Significant (p < 0.01) correlation between the eye dose and the kerma-area product was found for the first operator and radiographers, but not for other staff categories. The study revealed eye dose per procedure and eye dose normalised to patient dose indices for different staff categories and provided an input for radiation protection in electrophysiology procedures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.