Background: Osteocalcin (Oc), a serum marker of bone turnover, circulates in several forms. We developed an assay for intact human Oc and investigated its clinical features.
Methods: We generated goat antibodies and N- and C-terminal Oc. The former was used on solid phase (polystyrene beads), and the latter was used as the tracer in an IRMA.
Results: The assay was linear with no cross-reactivity to Oc(1–43), total imprecision (CV) of <10%, and recovery of 100% ± 10%. Assay values for intact Oc in EDTA plasma samples were unchanged at 18–25 °C for 6 h. Values for intact Oc in serum, EDTA plasma, and heparin plasma samples did not change after storage on ice for 8 h. Serum samples from patients with various conditions were stored at −70 or −135 °C for up to 5 years and yielded z-scores comparable to an Oc(1-43) IRMA for all conditions except for renal failure. In renal failure, the Oc(1–43) assay values were increased, whereas the intact assay values were in the reference interval.
Conclusion: Decreases in Oc assay values are inhibited by calcium chelation, and slowed by reduced temperatures. The described assay for intact Oc allows improved specificity for bone compared with an assay for Oc(1–43).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.