To test in vitro the hemolytic properties of pumps used in extracorporeal circulation, a system was made of polyvinyl chloride tubing. Hemolysis was expressed as the rate of accumulation of free hemoglobin in the plasma determined colorphotometrically. Human blood was recirculated with a single pump in the system. Several pumps—the Sigmamotor, the Leonards occlusive roller, the Hercik nonocclusive roller, the MacNeill-Collins and three modifications of the Dale-Schuster were tested under similar conditions. Some pumps in general clinical use produced as much as 19 mg/100 ml of free hemoglobin/hr. A modified Dale-Schuster pump and a Clowes membrane oxygenator also were tested in the system recirculating 800 ml of blood. The hourly production of free hemoglobin was only 3.2 mg/100 ml/hr. The concentration of free hemoglobin was kept constant by exchange transfusion with a modified Dale-Schuster pump and Clowes membrane oxygenator in the system. A formula used to calculate the required amount of blood to be exchanged is presented. Submitted on May 11, 1959
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.