In this randomized trial, cryoballoon ablation was noninferior to radiofrequency ablation with respect to efficacy for the treatment of patients with drug-refractory paroxysmal atrial fibrillation, and there was no significant difference between the two methods with regard to overall safety. (Funded by Medtronic; FIRE AND ICE ClinicalTrials.gov number, NCT01490814.).
Background-Contact force (CF) is a major determinant of lesion size and transmurality and has the potential to improve efficacy of atrial fibrillation ablation. This study sought to evaluate the safety and effectiveness of a novel irrigated radiofrequency ablation catheter that measures real-time CF in the treatment of patients with paroxysmal atrial fibrillation. Methods and Results-A total of 300 patients with symptomatic, drug-refractory, paroxysmal atrial fibrillation were enrolled in a prospective, multicenter, randomized, controlled trial and randomized to radiofrequency ablation with either a novel CF-sensing catheter or a non-CF catheter (control). The primary effectiveness end point consisted of acute electrical isolation of all pulmonary veins and freedom from recurrent symptomatic atrial arrhythmia off all antiarrhythmic drugs at 12 months. The primary safety end point included device-related serious adverse events. End points were powered to show noninferiority. All pulmonary veins were isolated in both groups. Effectiveness was achieved in 67.8% and 69.4% of subjects in the CF and control arms, respectively (absolute difference, −1.6%; lower limit of 1-sided 95% confidence interval, −10.7%; P=0.0073 for noninferiority). When the CF arm was stratified into optimal CF (≥90% ablations with ≥10 g) and nonoptimal CF groups, effectiveness was achieved in 75.9% versus 58.1%, respectively (P=0.018). The primary safety end point occurred in 1.97% and 1.40% of CF patients and control subjects, respectively (absolute difference, 0.57%; upper limit of 1-sided 95% confidence interval, 3.61%; P=0.0004 for noninferiority). Conclusions-The CF ablation catheter met the primary safety and effectiveness end points. Additionally, optimal CF was associated with improved effectiveness. Clinical Trial Registration-http://www.clinicaltrials.gov. Unique identifier: NCT01278953.(Circulation. 2015;132:907-915.
AimsThe primary safety and efficacy endpoints of the randomized FIRE AND ICE trial have recently demonstrated non-inferiority of cryoballoon vs. radiofrequency current (RFC) catheter ablation in patients with drug-refractory symptomatic paroxysmal atrial fibrillation (AF). The aim of the current study was to assess outcome parameters that are important for the daily clinical management of patients using key secondary analyses. Specifically, reinterventions, rehospitalizations, and quality-of-life were examined in this randomized trial of cryoballoon vs. RFC catheter ablation.Methods and resultsPatients (374 subjects in the cryoballoon group and 376 subjects in the RFC group) were evaluated in the modified intention-to-treat cohort. After the index ablation, log-rank testing over 1000 days of follow-up demonstrated that there were statistically significant differences in favour of cryoballoon ablation with respect to repeat ablations (11.8% cryoballoon vs. 17.6% RFC; P = 0.03), direct-current cardioversions (3.2% cryoballoon vs. 6.4% RFC; P = 0.04), all-cause rehospitalizations (32.6% cryoballoon vs. 41.5% RFC; P = 0.01), and cardiovascular rehospitalizations (23.8% cryoballoon vs. 35.9% RFC; P < 0.01). There were no statistical differences between groups in the quality-of-life surveys (both mental and physical) as measured by the Short Form-12 health survey and the EuroQol five-dimension questionnaire. There was an improvement in both mental and physical quality-of-life in all patients that began at 6 months after the index ablation and was maintained throughout the 30 months of follow-up.ConclusionPatients treated with cryoballoon as opposed to RFC ablation had significantly fewer repeat ablations, direct-current cardioversions, all-cause rehospitalizations, and cardiovascular rehospitalizations during follow-up. Both patient groups improved in quality-of-life scores after AF ablation.Clinical trial registrationClinicalTrials.gov identifier: NCT01490814.
Non-invasive mapping identifies biatrial drivers that are critical in PsAF. This is validated by successful AF termination in the majority of patients treated in centres with no experience of the system. Ablation targeting these drivers results in favourable AF-free survival at 1 year, albeit with a significant rate of AT recurrence requiring further management.
Pulmonary vein isolation using CF-guided RF and second-generation CB leads to comparable single-procedure arrhythmia-free survival at up to 18 months with similar overall complication rate.
Background:
The FIRE AND ICE trial assessed efficacy and safety of pulmonary vein (PV) isolation using cryoballoon versus radiofrequency current (RFC) ablation in patients with drug refractory, symptomatic, paroxysmal atrial fibrillation (AF). The purpose of the current study was to assess index lesion durability as well as reablation strategy and outcomes in trial patients undergoing a reablation procedure.
Methods:
Patients with reablation procedures during FIRE AND ICE were retrospectively consented and enrolled at 13 trial centers. The first reablation for each patient was included in the analysis. Documented arrhythmias before reablation, number and location of reconnected PVs, lesions created during reablations, procedural characteristics, and acute as well as long-term outcomes were assessed.
Results:
Eighty-nine (36 cryoballoon and 53 RFC) patients were included in this study. Paroxysmal atrial fibrillation was the predominant recurrent arrhythmia (69%) before reablation. Reablations occurred at a median of 173 and 182 days (
P
=0.54) in the cryoballoon and RFC cohorts, respectively. The number of reconnected PVs was significantly higher in the RFC than the cryoballoon group (2.1±1.4 versus 1.4±1.1;
P
=0.010), which was driven by significantly more reconnected left superior PVs and markedly more reconnected right superior PVs. The number of (predominantly RFC) lesions applied during reablation was significantly greater in patients originally treated with RFC (3.3±1.3 versus 2.5±1.5;
P
=0.015) with no difference in overall acute success (
P
=0.70). After reablation, no differences in procedure-related rehospitalization or antiarrhythmic drug utilization were observed between cohorts.
Conclusions:
At reablation, patients originally treated with the cryoballoon had significantly fewer reconnected PVs, which may reflect RFC catheter instability in certain left atrial regions, and thus required fewer lesions for reablation success. Repeat ablations were predominantly performed with RFC and resulted in similar acute success, duration of hospitalization, and antiarrhythmic drug prescription between the study cohorts.
Clinical Trial Registration:
URL:
https://www.clinicaltrials.gov
. Unique identifier: NCT03314753.
Our findings suggest that mid-term outcomes of cryoballoon ablation for paroxysmal AF ablation are similar to those of radiofrequency, regardless of PV anatomy. The presence of anatomical variants of PVs should not discourage the referral of patients with paroxysmal AF for cryoballoon ablation.
The incidence of sustained VT remains relatively low in the first year after CRT-P implantation, but when present appears closely associated with short-term adverse outcomes, especially SCD. This emphasizes the possible value of remote monitoring to detect high-risk patients for urgent upgrading.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.