Our group has proposed the concept of subharmonic aided pressure estimation (SHAPE) utilizing microbubble-based ultrasound contrast agent signals for the noninvasive estimation of hydrostatic blood pressures. An experimental system for in vitro SHAPE was constructed based on two single-element transducers assembled confocally at a 60 degree angle to each other. Changes in the first, second and subharmonic amplitudes of five different ultrasound contrast agents were measured in vitro at static hydrostatic pressures from 0-186 mmHg, acoustic pressures from 0.35-0.60 MPa peak-to-peak and frequencies of 2.5-6.6 MHz. The most sensitive agent and optimal parameters for SHAPE were determined using linear regression analysis and implemented on a Logiq 9 scanner (GE Healthcare, Milwaukee, WI). This implementation of SHAPE was then tested under dynamic-flow conditions and compared to pressure-catheter measurements. Over the pressure range studied, the first and second harmonic amplitudes reduced approximately 2 dB for all contrast agents. Over the same pressure range, the subharmonic amplitudes decreased by 9-14 dB and excellent linear regressions were achieved with the hydrostatic pressure variations (r = 0.98, p < 0.001). Optimal sensitivity was achieved at a transmit frequency of 2.5 MHz and acoustic pressure of 0.35 MPa using Sonazoid (GE Healthcare, Oslo, Norway). A Logiq 9 scanner was modified to implement SHAPE on a convex transducer with a frequency range from 1.5-4.5 MHz and acoustic pressures from 0-3.34 MPa. Results matched the pressure catheter (r2 = 0.87). In conclusion, subharmonic contrast signals are a good indicator of hydrostatic pressure. Out of the five ultrasound contrast agents tested, Sonazoid was the most sensitive for subharmonic pressure estimation. Real-time SHAPE has been implemented on a commercial scanner and offers the possibility of allowing pressures in the heart and elsewhere to be obtained noninvasively.
Purpose:To compare subharmonic aided pressure estimation (SHAPE) with pressure catheter-based measurements in human patients with chronic liver disease undergoing transjugular liver biopsy. Materials and Methods:This HIPAA-compliant study had U.S. Food and Drug Administration and institutional review board approval, and written informed consent was obtained from all participants. Forty-five patients completed this study between December 2010 and December 2011. A clinical ultrasonography (US) scanner was modified to obtain SHAPE data. After transjugular liver biopsy with pressure measurements as part of the standard of care, 45 patients received an infusion of a microbubble US contrast agent and saline. During infusion, SHAPE data were collected from a portal and hepatic vein and were compared with invasive measurements. Correlations between data sets were determined by using the Pearson correlation coefficient, and statistical significance between groups was determined by using the Student t test. Results:-The 45 study patients included 27 men and 18 women (age range, 19-71 years; average age, 55.8 years). The SHAPE gradient between the portal and hepatic veins was in good overall agreement with the hepatic venous pressure gradient (HVPG) (R = 0.82). Patients at increased risk for variceal hemorrhage (HVPG > 12 mm Hg) had a significantly higher mean subharmonic gradient than patients with lower HVPGs (1.93 dB 6 0.61 [standard deviation] vs 21.47 dB 6 0.29, P , .001), with a sensitivity of 100% and a specificity of 81%, indicating that SHAPE may be a useful tool for the diagnosis of clinically important portal hypertension. Conclusion:Preliminary results show SHAPE to be an accurate noninvasive technique for estimating portal hypertension.q RSNA, 2013 Supplemental material: http://radiology.rsna.org/lookup /suppl
The purpose of this study was to develop and investigate a new noninvasive approach to quantify left ventricular (LV) pressures using subharmonic emissions from microbubbles. A Sonix RP ultrasound scanner with PA4-2 phased array transducer was used in pulse inversion grayscale mode. Unprocessed radiofrequency data were obtained for 5 seconds (n=3) with pulsed wave Doppler from the aorta and/or LV of 4 canines during Sonazoid infusion. Simultaneous pressure measurements were obtained using Millar manometer. Subharmonic data (in dB) were extracted and processed. The resulting calibration factor (mmHg/dB), from the aorta, was used to estimate LV pressures. Errors ranged from 0.19 to 2.50 mmHg when estimating these pressures using the aortic calibration factor from the respective canines; but were considerably higher (0.64-8.98 mmHg) when a mean aortic calibration factor was used. In conclusion, subharmonic emissions from ultrasound contrast agents have the potential to noninvasively monitor LV pressures.
The feasibility of using subharmonic aided pressure estimation (SHAPE) to noninvasively estimate interstitial fluid pressure (IFP) was studied. In vitro, radiofrequency signals, from 0.2 ml/l of Definity (Lantheus Medical Imaging, N Billerica, MA) were acquired within a water-tank with a Sonix RP ultrasound scanner (Ultrasonix, Richmond, BC, Canada; fT/R=6.7/3.35 MHz and fT/R =10/5 MHz) and the subharmonic amplitudes of the signals were compared over 0–50 mmHg. In vivo, five swine with naturally occurring melanomas were studied. Subharmonic signals were acquired from tumours and surrounding tissue during infusion of Definity and compared to needle-based pressure measurements. Both in vitro and in vivo, an inverse linear relationship between hydrostatic pressure and subharmonic amplitude was observed with r2=0.63–0.95; p<0.05, maximum amplitude drop 11.36 dB at 10 MHz and −8 dB, and r2 as high as 0.97; p<0.02 (10 MHz and −4/−8 dB most promising), respectively, indicating that SHAPE may be useful in monitoring IFP.
Objective To demonstrate the feasibility of simultaneous dual fundamental grayscale and subharmonic imaging on a modified commercial scanner. Motivation The ability to generate signals at half the insonation frequency is exclusive to ultrasound contrast agents (UCA). Thus, subharmonic imaging (SHI; transmitting at f0 and receiving at f0/2) provides improved visualization of UCA within the vasculature via suppression of the surrounding tissue echoes. While this capability has proven useful in a variety of clinical applications, the SHI suppression of surrounding tissue landmarks (which are needed for sonographic navigation) also limits it use as a primary imaging modality. In this paper we present results using a commercial ultrasound scanner modified to allow imaging in both grayscale (f0 = 4.0 MHz) and SHI (f0 = 2.5 MHz, f0/2 = 1.25 MHz) modes in real time. Methods A Logiq 9 ultrasound scanner (GE Healthcare, Milwaukee, WI) with a 4C curvilinear probe was modified to provide this capability. Four commercially available UCA (Definity, Lantheus Medical Imaging, North Billerica, MA; Optison, GE Healthcare, Princeton, NJ; SonoVue Bracco Imaging, Milan, Italy; and Sonazoid GE Healthcare, Oslo, Norway) were all investigated in vitro over an acoustic output range of 3.34 MPa. In vivo the subharmonic response of Sonazoid (GE Healthcare, Oslo, Norway) was investigated in the portal veins of 4 canines (open abdominal cavity) and 4 patients with suspected portal hypertension. Results In vitro, the four UCA showed an average maximum subharmonic amplitude of 44.1 ± 5.4 dB above the noise floor with a maximum subharmonic amplitude of 48.6 ± 1.6 dB provided by Sonazoid. The average in vivo maximum signal above the noise floor from Sonazoid was 20.8 ± 2.3 dB in canines and 33.9 ± 5.2 dB in humans. Subharmonic amplitude as a function of acoustic output in both groups matched the S-curve behavior if the agent observed in vitro. The dual grayscale imaging provided easier sonographic navigation while the degree of tissue suppression in SHI mode varied greatly on a case by case basis. Conclusions These results demonstrate the feasibility of dual grayscale and SHI on a modified commercial scanner. The ability to simultaneously visualize both imaging modes in real time should improve the applicability of SHI as a future primary clinical imaging modality.
The efficacy of using subharmonic emissions from Sonazoid microbubbles (GE Healthcare, Oslo, Norway) to track portal vein pressures and pressure changes was investigated in 14 canines using either slow- or high-flow models of portal hypertension (PH). A modified Logiq 9 scanner (GE Healthcare, Milwaukee, WI) operating in subharmonic mode (ftransmit:2.5MHz, freceive:1.25MHz) was used to collect RF data at 10-40% incident acoustic power levels with 2-4 transmit cycles (in triplicate), before and after inducing PH. A pressure catheter (Millar Instruments, Inc., Houston, TX) provided reference portal vein pressures. At optimum insonification, subharmonic signal amplitude changes correlated with portal vein pressure changes; r ranged from -0.82 to -0.94 and from -0.70 to -0.73 for PH models considered separately or together, respectively. The subharmonic signal amplitudes correlated with absolute portal vein pressures (r: -0.71 to -0.79). Statistically significant differences between subharmonic amplitudes, before and after inducing PH, were noted (p≤0.01). Portal vein pressures estimated using SHAPE did not reveal significant differences (p>0.05) with respect to the pressures obtained using the Millar pressure catheter. Subharmonic aided pressure estimation may be useful clinically for portal vein pressure monitoring.
Parametric maps showing perfusion of contrast media can be useful tools for characterizing lesions in breast tissue. In this study we show the feasibility of parametric subharmonic imaging (SHI), which allows imaging of a vascular marker (the ultrasound contrast agent) while providing near complete tissue suppression. Digital SHI clips of 16 breast lesions from 14 women were acquired. Patients were scanned using a modified LOGIQ 9 scanner (GE Healthcare, Waukesha, WI) transmitting/receiving at 4.4/2.2 MHz. Using motion-compensated cumulative maximum intensity (CMI) sequences, parametric maps were generated for each lesion showing the time to peak (TTP), estimated perfusion (EP), and area under the time-intensity curve (AUC). Findings were grouped and compared according to biopsy results as benign lesions (n = 12, including 5 fibroadenomas and 3 cysts) and carcinomas (n = 4). For each lesion CMI, TTP, EP, and AUC parametric images were generated. No significant variations were detected with CMI (P = .80), TTP (P = .35), or AUC (P = .65). A statistically significant variation was detected for the average pixel EP (P = .002). Especially, differences were seen between carcinoma and benign lesions (mean ± SD, 0.10 ± 0.03 versus 0.05 ± 0.02 intensity units [IU]/s; P = .0014) and between carcinoma and fibroadenoma (0.10 ± 0.03 versus 0.04 ± 0.01 IU/s; P = .0044), whereas differences between carcinomas and cysts were found to be nonsignificant. In conclusion, a parametric imaging method for characterization of breast lesions using the high contrast to tissue signal provided by SHI has been developed. While the preliminary sample size was limited, results show potential for breast lesion characterization based on perfusion flow parameters.
The purpose of this study was to develop and validate a noninvasive pressure estimation technique based on subharmonic emissions from a commercially available ultrasound contrast agent and scanner, unlike other studies that have either adopted a single-element transducer approach and/ or use of in-house contrast agents. Ambient pressures were varied in a closed-loop flow system between 0 and 120 mmHg and were recorded by a solid-state pressure catheter as the reference standard. Simultaneously, the ultrasound scanner was operated in pulse inversion mode transmitting at 2.5 MHz, and the unprocessed RF data were captured at different incident acoustic pressures (from 76 to 897 kPa). The subharmonic data for each pulse were extracted using band-pass filtering with averaging, and subsequently processed to eliminate noise. The incident acoustic pressure most sensitive to ambient pressure fluctuations was determined, and then the ambient pressure was tracked over 20 s. In vivo validation of this technique was performed in the left ventricle (LV) of 2 canines. In vitro, the subharmonic signal could track ambient pressure values with r(2) = 0.922 (p < 0.001), whereas in vivo, the subharmonic signal tracked the LV pressures with r(2) > 0.790 (p < 0.001) showing a maximum error of 2.84 mmHg compared with the reference standard. In conclusion, a subharmonic ultrasound-based pressure estimation technique, which can accurately track left ventricular pressures, has been established.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.