In this nationwide snapshot survey of cardioversion procedures in Germany, approximately 11% of patients presenting for elective electrical cardioversion were pacemaker or ICD carriers. Cardioversion protocols in these patients are heterogeneous throughout centers and mostly not in accordance with recommendation of the German Cardiac Society. Complications associated with external electrical cardioversion are rare. Controlled trials and large registries are necessary to provide evidence for future recommendations.
Background
Current implantable cardioverter-defibrillator (ICD) guidelines do not impose age limitations for ICD implantation (IMPL) and generator exchange (GE); however, patients (pts) should be expected to survive for 1 year. With higher age, comorbidity and mortality due to non-sudden cardiac death increase. Thus, the benefit of ICD therapy in elderly pts remains unclear. Mortality after ICD IMPL or GE in pts ≥ 75 years was assessed.
Methods
Consecutive pts aged ≥ 75 years with ICD IMPL or GE at the University Hospital Cologne, Germany, between 01/2013 and 12/2017 were included in this retrospective analysis.
Results
Of 418 pts, 82 (20%) fulfilled the inclusion criteria; in 70 (55 = IMPL, 79%, 15 = GE, 21%) follow-up (FU) was available. The median FU was 3.1 years. During FU, 40 pts (57%) died (29/55 [53%] IMPL; 11/15 [73%] GE). Mean survival after surgery was 561 ± 462 days. The 1‑year mortality rate was 19/70 (27%) overall, 9/52 (17%) in pts ≥ 75 and 10/18 (56%) in pts ≥ 80 years. Deceased pts were more likely to suffer from chronic renal failure (85% vs. 53%, p = 0.004) and peripheral artery disease (18% vs. 0%, p = 0.02). During FU, seven pts experienced ICD shocks (four appropriate, three inappropriate). In primary prevention (n = 35) mortality was 46% and four pts experienced ICD therapies (two adequate); in secondary prevention (n = 35) mortality was 69% (p = 0.053) with three ICD therapies (two adequate).
Conclusion
Mortality in ICD pts aged ≥ 80 years was 56% at 1 and 72% at 2 years in this retrospective analysis. The decision to implant an ICD in elderly pts should be made carefully and individually.
AF ablation can be performed safely, with high acute success rates using RAN. RNS results in less fluoroscopy burden and shorter procedure durations. Compared to SMN, a reduced fluoroscopy burden, prolonged procedure and ablation duration were observed using RAN. Overall, the number of RAN procedures is small suggesting low impact on clinical routine of AF ablation.
Background
Omnipolar mapping (OT) is a novel tool to acquire omnipolar signals for electro-anatomical mapping, displaying true voltage and real-time wavefront direction and speed independent of catheter orientation. The aim was to analyze previously performed left atrial (LA) and left ventricular (LV) maps for differences using automated OT vs. standard bipolar settings (SD) and HD wave (HDW) algorithm.
Methods
Previously obtained SD and HDW maps of the LA and LV using a 16-electrode, grid-shaped catheter were retrospectively analyzed by applying automated OT, comparing voltage, point density, pulmonary vein (PV) gaps, and LV scar area.
Results
In this analysis, 135 maps of 45 consecutive patients (30 treated for LA, 15 for LV arrhythmia) were included. Atrial maps revealed significantly higher point densities using OT (21471) vs. SD (6682) or HDW (12189, p < 0.001). Mean voltage was significantly higher using OT (0.75 mV) vs. SD (0.61 mV) or HDW (0.64 mV, p < 0.001). OT maps detected significantly more PV gaps per patient vs. SD (4 vs. 2), p = 0.001.
In LV maps, OT revealed significantly higher point densities (25951) vs. SD (8582) and HDW (17071), p < 0.001. Mean voltage was significantly higher for OT (1.49 mV) vs. SD (1.19 mV) and HDW (1.2 mV), p < 0.001. Detected scar area was significantly smaller using OT (25.3%) vs. SD (33.9%, p < 0.001).
Conclusion
OT mapping leads to significantly different substrate display, map density, voltage, detection of PV gaps, and scar size, compared to SD and HDW in LA and LV procedures. Successful CA might be facilitated due to true HD maps.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.