In this work, the optimum homogeneous phantom size for an equivalent whole-body electromagnetic (EM) modeling is calculated. This will enable the simple characterization of plane wave EM attenuation and far-field link budgets in Active Medical Implant (AMI) applications in the core region of the body for Industrial, Scientific, Medical and MedRadio frequency bands. A computational analysis is done to determine the optimum size in which a minimum phantom size reliably represents a whole-body situation for the corresponding frequency of operation, saving computer and laboratory resources. After the definition of a converge criterion, the computed minimum phantom size for subcutaneous applications, 0–10 mm insertion depth, is 355 × 160 × 255 mm3 for 402 MHz and 868 MHz and a cube with a side of 100 mm and 50 mm for 2.45 GHz and 5.8 GHz, respectively. For deep AMI applications, 10–50 mm insertion depth, the dimensions are 355 × 260 × 255 mm3 for 402 MHz and 868 MHz, and a cube with a side of 200 mm and 150 mm for 2.45 GHz and 5.8 GHz, respectively. A significant reduction in both computational and manufacturing resources for phantom development is thereby achieved. The verification of the model is performed by field measurements in phantoms made by aqueous solutions with sugar.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.