Implantation of the LV lead for biventricular pacing can be challenging, time consuming, and often requires extensive fluoroscopy time. A conventional diagnostic 5 Fr left Amplatz catheter was used to cannulate the coronary sinus in 15 consecutive patients undergoing implantation of a biventricular pacemaker. When the coronary sinus was cannulated, the proximal end of the Amplatz catheter was cut and the coronary sinus sheath was passed over the Amplatz catheter that was then removed. Coronary sinus cannulation was achieved in all 15 patients with a mean fluoroscopy time of 3.34 +/- 1.9 minutes. Subsequent implantation of a biventricular pacemaker was successful and free of complications in all the 15 patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.