The multireflection ion traps with isochronous properties offer a Lot of opportunities for time-of-flight mass spectrometry by elongation of the ion path, thus preserving the compact dimensions of an instrument. We have built and tested a two-mirror linear trap that provides at least 80,000 mass-resolving power. Although the mass resolution appears promising, there are substantial limitations that arise from Coulomb interactions of the trapped ions. Among these, the mutual repulsion of ions with same or close mass-to-charge ratios appears dominant, resulting in counterintuitive motion synchronization. The self-bunching and coalescence effects are also examined by numerical simulation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.