Purpose
To determine patient factors and eye conditions associated with artifacts in Spectralis optical coherence tomography (OCT) retinal nerve fiber layer (RNFL) scans.
Design
Retrospective cross-sectional study.
Methods
The prevalence of twelve artifact types were described in this review of 2313 eye scans from 1188 patients who underwent a complete eye exam with Spectralis OCT scanning during the period of September 2009 to July 2013. Generalized estimating equations model was utilized to analyze associations between increased artifact prevalence and 10 patient characteristics, which included age, sex, race, visual acuity, refractive error, astigmatism, cataract status, glaucoma staging, visual field reliability, and glaucoma diagnosis.
Results
A total of 1070 or 46.3% of the 2313 eye scans had at least one artifact. De-centration error was the most common artifact (27.8%), followed by posterior vitreous detachment artifacts (14.4%). Visual acuity of less than 20/40 (p<0.0001), presence of moderate to severe cataracts (p<0.0001), advanced stage of glaucoma (p<0.0001), and a diagnosis of open angle glaucoma (p=0.0003) were associated with increased prevalence of artifacts.
Conclusions
Clinicians should first assess scans for artifacts before making therapeutic decisions based on RNFL thickness measurements.
PurposeTo describe spectral-domain optical coherence tomography (OCT) methods for quantifying neuroretinal rim tissue in glaucoma and to compare these methods to the traditional retinal nerve fiber layer thickness diagnostic parameter.MethodsNeuroretinal rim parameters derived from three-dimensional (3D) volume scans were compared with the two-dimensional (2D) Spectralis retinal nerve fiber layer (RNFL) thickness scans for diagnostic capability. This study analyzed one eye per patient of 104 glaucoma patients and 58 healthy subjects. The shortest distances between the cup surface and the OCT-based disc margin were automatically calculated to determine the thickness and area of the minimum distance band (MDB) neuroretinal rim parameter. Traditional 150-μm reference surface–based rim parameters (volume, area, and thickness) were also calculated. The diagnostic capabilities of these five parameters were compared with RNFL thickness using the area under the receiver operating characteristic (AUROC) curves.ResultsThe MDB thickness had significantly higher diagnostic capability than the RNFL thickness in the nasal (0.913 vs. 0.818, P = 0.004) and temporal (0.922 vs. 0.858, P = 0.026) quadrants and the inferonasal (0.950 vs. 0.897, P = 0.011) and superonasal (0.933 vs. 0.868, P = 0.012) sectors. The MDB area and the three neuroretinal rim parameters based on the 150-μm reference surface had diagnostic capabilities similar to RNFL thickness.ConclusionsThe 3D MDB thickness had a high diagnostic capability for glaucoma and may be of significant clinical utility. It had higher diagnostic capability than the RNFL thickness in the nasal and temporal quadrants and the inferonasal and superonasal sectors.
Purpose
To determine the diagnostic capability of spectral domain optical coherence tomography (SD-OCT) peripapillary retinal thickness (RT) measurements from 3-dimensional (3D) volume scans for primary open angle glaucoma (POAG).
Design
Cross-sectional study.
Methods
Setting
Institutional
Study population
156 patients (89 POAG and 67 normal subjects)
Observation procedures
One eye of each subject was included. SD-OCT peripapillary RT values from 3D volume scans were calculated for four quadrants of three different sized annuli. Peripapillary retinal nerve fiber layer (RNFL) thickness values were also determined.
Main outcome measures
Area under the receiver operating characteristic curve (AUROC) values, sensitivity, specificity, positive and negative predictive values, and positive and negative likelihood ratios.
Results
The top five RT AUROCs for all glaucoma patients and for a subset of early glaucoma patients were for the inferior quadrant of outer circumpapillary annulus of circular grid (OCA) 1 (0.959, 0.939), inferior quadrant of OCA2 (0.945, 0.921), superior quadrant of OCA1 (0.890, 0.811), inferior quadrant of OCA3 (0.887, 0.854), and superior quadrant of OCA2 (0.879, 0.807). Smaller RT annuli OCA1 and OCA2 consistently showed better diagnostic performance than the larger RT annulus OCA3. For both RNFL and RT measurements, best AUROC values were found for inferior RT OCA1 and OCA2, followed by inferior and overall RNFL thickness.
Conclusion
Peripapillary RT measurements from 3D volume scans showed excellent diagnostic performance for detecting both glaucoma and early glaucoma patients. Peripapillary RT values have the same or better diagnostic capability compared to peripapillary RNFL thickness measurements, while also having fewer algorithm errors.
The thickness of anterior wall structures and preocular tear film of patients with moderate to high myopia are not statistically different compared with those of healthy controls.
The average RNFL thickness in the migraine patients was found to be thinner than that in the control group. In addition, we found a negative weak correlation between length of migraine history and the average RNFL thickness, supporting the possible association between these pathologies.
Purpose
To compare the diagnostic capability of 3-dimensional (3D) neuroretinal rim parameters with existing 2-dimensional (2D) neuroretinal and retinal nerve fiber layer (RNFL) thickness rim parameters using spectral domain optical coherence tomography (SD-OCT) volume scans
Materials and Methods
Design
Institutional prospective pilot study.
Study population
65 subjects (35 open angle glaucoma patients, 30 normal patients).
Observation procedures
One eye of each subject was included. SD-OCT was used to obtain 2D retinal nerve fiber layer (RNFL) thickness values and five neuroretinal rim parameters [i.e. 3D minimum distance band (MDB) thickness, 3D Bruch’s membrane opening-minimum rim width (BMO-MRW), 3D rim volume, 2D rim area, and 2D rim thickness].
Main outcome measures
Area under the receiver operating characteristic (AUROC) curve values, sensitivity, specificity.
Results
Comparing all 3D with all 2D parameters, 3D rim parameters (MDB, BMO-MRW, rim volume) generally had higher AUROC curve values (range 0.770–0.946) compared to 2D parameters (RNFL thickness, rim area, rim thickness; range 0.678–0.911). For global region analyses, all 3D rim parameters (BMO-MRW, rim volume, MDB) were equal to or better than 2D parameters (RNFL thickness, rim area, rim thickness; p-values from 0.023–1.0). Among the three 3D rim parameters (MDB, BMO-MRW, and rim volume), there were no significant differences in diagnostic capability (false discovery rate > 0.05 at 95% specificity).
Conclusion
3D neuroretinal rim parameters (MDB, BMO-MRW, and rim volume) demonstrated better diagnostic capability for primary and secondary open angle glaucomas compared to 2D neuroretinal parameters (rim area, rim thickness). Compared to 2D RNFL thickness, 3D neuroretinal rim parameters have the same or better diagnostic capability.
Intereye RNFL asymmetry may be a useful clinical OCT measurement to provide quantitative assessment of early glaucomatous damage. Newly developed algorithms for intereye RNFL asymmetry may improve the ability to detect glaucoma.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.