Background
In this label-free bioassay, an electrochemiluminescence (ECL) immunosensor was developed for the quantification of breast cancer using HER-2 protein as a metastatic biomarker.
Method
For this purpose, the ECL emitter, [Ru(bpy)3]2+, was embedded into biocompatible chitosan (CS) polymer. The prepared bio-composite offered high ECL reading due to the depletion of human epidermal growth factor receptor 2 (HER-2) protein. Reduced graphene oxide (rGO) was used as substrate to increase signal stability and achieve greater sensitivity. For this, rGO was initially placed electrochemically on the glassy carbon electrode (GCE) surface by cyclic voltammetry (CV) technique. Next, the prepared CS/[Ru(bpy)3]2+ biopolymer solution was coated on a drop of the modified electrode such that the amine groups of CS and the carboxylic groups of rGO could covalently interact. Using EDC/NHS chemistry, monoclonal antibodies (Abs) of HER-2 were linked to CS/[Ru(bpy)3]2+/rGO/GCE via amide bonds between the carboxylic groups of Ab molecules and amine groups of CS. The electrochemical behavior of the electrode was studied using different electrochemical techniques such as electrochemical impedance spectroscopy (EIS), differential pulse voltammetry (DPV) and square wave voltammetry (SWV) and also ECL tests.
Results
After passing all optimization steps, the lower limit of detection (LLOQ) and linear dynamic range (LDR) of HER-2 protein were practically obtained as 1 fM and 1 fM to 1 nM, individually. Importantly, the within and between laboratory precisions were performed and the suitable relative standard deviations (RSDs) were recorded as 3.1 and 3.5%, respectively.
Conclusions
As a proof of concept, the designed immunosensor was desirably applied for the quantification of HER-2 protein in breast cancer suffering patients. As a result, the designed ECL-based immunosensor has the capability of being used as a conventional test method in biomedical laboratories for early detection of HER-2 protein in biological fluids.
Graphic Abstract
Biopolymer films have drawn growing demand for their application in the point of care domain owing to their biocompatibility, eco-friendly, and eligibility for in vivo analyses. However, their poor conductivity restricts their sensitivity in diagnostics. For high-quality electrochemical biosensor monitoring, two vital factors to be greatly paid attention are the effective merge of amplification modifiers with transducing surface and the superior linking across the recognition interface. Here, we introduce an enzyme-free electrochemical biosensor based on electrosynthesized biocompatible WO3/poly glutamic acid nano-biocomposites to address the hardships specific to the analysis of circulating proteins clinical samples. In addition to its green synthesis route, the poor tendency of both components of the prepared nano-biocomposite to amine groups makes it excellent working in untreated biological samples with high contents of proteins. Several electrochemical and morphological investigations (SEM, EDX, and dot mapping) were fulfilled to gain a reliable and trustful standpoint of the framework. By using this nanobiosensor, the concentration of HER-2 was detectable as low as 1 fg mL−1 with a wide linear response between 1 ng mL−1 and 1 fg mL−1. Meanwhile, the protocol depicted ideal specificity, stability, and reproducibility for the detection of HER-2 protein in untreated serum samples of breast cancer patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.