We demonstrated that not only the regional limits of agreement are too wide to allow data of the two montages to be used interchangeably but also the variation is a function of anesthetic depth. However, keeping in mind a relatively consistent BIS-Vista -10 bias before induction and +10 bias during anesthesia maintenance with limits of agreement of approximately +/-11 BIS units, approximately double the clinically acceptable less than 10 BIS units level of agreement, BIS-Vista off-label occipital montage might be helpful in following a trend of propofol-remifentanil anesthesia in individual cases where frontal access is particularly difficult.
Although we demonstrated good correlation and high discriminative power of the BIS-Vista monitor in depicting a MCBFV decline that could serve as indicator of decline in cerebral activity, BIS-Vista cannot be considered a reliable indicator of cerebral ischemia/hypoperfusion that could replace transcranial Doppler monitoring to determine whether a shunt is to be placed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.