Background:In the emerging field of robot-assisted spine surgery, radiographic evaluation of pedicle screw accuracy in the surgical setting is of high interest. Advances in medical imaging have improved the accuracy of pedicle screw placement, from fluoroscopy-guided to computer-aided navigation.
Methods:A retrospective, institutional review board-exempt review of the first 106 navigated robot-assisted spine surgery cases was performed. Radiographic evaluation of preoperative and postoperative computerized tomography (CT) scans were collected.Results: In the first 106 cases, 630 lumbosacral pedicle screws were placed. Thirty screws were placed in five patients without the robot because of surgeon discretion.Of the 600 pedicle screws inserted by navigated robotic guidance, only 1.5% (9/600) were repositioned intraoperatively.
Conclusion:This study demonstrated a high level of accuracy (98.2%) in terms of grade A or B pedicle screw breach scores in the clinical use of navigated, robotassisted surgery in its first 101 cases.
Robotic assistance with integrated navigation is an area of high interest for improving the accuracy of minimally invasive pedicle screw placement. This study analyzes the accuracy of pedicle screw placement between an attending spine surgeon and a resident by comparing the left and right sides of the first 101 consecutive cases using navigated robotic assistance in a private practice clinical setting. A retrospective, Institutional Review Board-exempt review of the first 106 navigated robot-assisted spine surgery cases was performed. One attending spine surgeon and one resident performed pedicle screw placement consistently on either the left or right side (researchers were blinded). A CT-based Gertzbein and Robbins system (GRS) was used to classify pedicle screw accuracy, with grade A or B considered accurate. There were 630 consecutive lumbosacral pedicle screws placed. Thirty screws (5 patients) were placed without the robot due to surgeon discretion. Of the 600 pedicle screws inserted by navigated robotic guidance (101 patients), only 1.5% (9/600) were repositioned intraoperatively. Based on the GRS CT-based grading of pedicle breach, 98.67% (296/300) of left-side screws were graded A or B, 1.3% (4/300) were graded C, and 0% (0/300) were graded D. For the right-side screws, 97.67% (293/300) were graded A or B, 1.67% (5/300) were graded C, and 0.66% (2/300) were graded D. This study demonstrated a high level of accuracy (based on GRS) with no significant differences between the left- and right-side pedicle screw placements (98.67% vs. 97.67%, respectively) in the clinical use of navigated, robot-assisted surgery.
A 63-year-old man presented with progressive lower-extremity paresis over the previous 3 months. He had been unable to support himself to ambulate over the previous 3 weeks. The patient was found to have metastatic renal cell carcinoma to T11. He underwent robotic assisted percutaneous placement of T9–L1 pedicle screws and open T11 laminectomy/debulking of tumor. This approach allowed for decreased blood loss without the need for preoperative embolization. After decompression and placement of the hardware, the patient regained the ability to ambulate on his own.The video can be found here: https://youtu.be/BTFQezhZlB4.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.