A carbon paste electrode (CPE) modified with gold nanoparticles was used as a selective electrochemical sensor for the determination of tryptophan (Trp), uric acid (UA) and ascorbic acid (AA). Cyclic voltammetric, chronoamperometric, and differential pulse voltammetric methods were used to investigate the electrocatalytic oxidation of Trp in aqueous solution. The Au nanoparticles modified carbon paste electrode (AuNPs/CPE) showed very efficient electrocatalytic activity for anodic oxidation of Trp, UA, and AA. The linear calibration range for Trp in the presence of 100 mM AA and 50 mM UA was 6-200 mM, with a detection limit of 0.65 mM (S/N ¼ 3). A linear relationship was found for UA in the range of 6-180 mM containing 100 mM AA and 25 mM Trp, with a detection limit of 0.71 mM. The linear calibration range for AA in the presence of 50 mM UA and 50 mM Trp was 20-250 mM, with a detection limit of 2.05 mM. The diffusion coefficient (D) for the oxidation of Trp at the modified electrode was calculated as 2.82 Â 10 À5 cm 2 s À1 . The modified electrode showed good sensitivity and selectivity, and was employed for the determination of Trp, UA, and AA in biological and pharmaceutical samples.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.