In ambulatory surgery, anesthetic drugs must be administered at a suitable rate to prevent adverse reactions after discharge from the hospital. To realize more appropriate anesthesia, we have developed a hypnosis control system, which administers propofol as an anesthetic drug to regulate the bispectral index (BIS), an electroencephalography (EEG)-derived index reflecting the hypnosis of a patient. This system consists of three functions: 1) a feedback controller using a model-predictive control method, which can adequately accommodate the effects of time delays; 2) a parameter estimation function of individual differences; and 3) a risk control function for preventing undesirable states such as drug overinfusion or intraoperative arousal. With the approval of the ethics committee of our institute, 79 clinical trials took place since July 2002. The results show that our system can reduce the total amount of propofol infusion and maintain the BIS more accurately than anesthesiologist's manual adjustment.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.