Biomagnetic measurement performed outside a magnetically shielded room is subject to distortion by strong magnetic fields. Reducing such disturbances can enhance and stabilize biomagnetic measurement conditions in the absence of passive shielding. We have developed an active magnetic shielding system that produces both homogeneous and spatial gradient magnetic fields. The system is composed of anisotropic magnetoresistive sensors, a digital signal processor controller and two different coil systems. In order to improve the measurement environment for a first-order gradient coil SQUID system, the disturbing vertical magnetic fields and vertical field gradients are reduced, thus achieving a shielding factor of approximately 6 at 100 Hz. Our system provides a more flexible and less costly alternative to magnetically shielded rooms.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.