A double-beam laser fluorimeter, using a single optical fiber to guide the lights, was constructed for in situ and on-line monitoring of NADH concentration [( NADH]) from normally blood-perfused living tissues. The device was tested on an isolated blood-perfused rat heart system to determine the most efficient reference wavelength for the compensation of the hemodynamic artifact induced by blood circulation in the tissues on the fluorescence measure; 586 nm was found to be an accurate reference wavelength, and a mathematical relationship was established that allowed the digital treatment of the measured fluorescence to give a signal (compensated fluorescence) that varied only with [NADH] in the volume of tissue investigated.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations鈥揷itations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.