Blood in the splanchnic vasculature can be transferred to the extremities. We quantified such blood shifts in normal subjects by measuring trunk volume by optoelectronic plethysmography, simultaneously with changes in body volume by whole body plethysmography during contractions of the diaphragm and abdominal muscles. Trunk volume changes with blood shifts, but body volume does not so that the blood volume shifted between trunk and extremities (Vbs) is the difference between changes in trunk and body volume. This is so because both trunk and body volume change identically with breathing and gas expansion or compression. During tidal breathing Vbs was 50–75 ml with an ejection fraction of 4–6% and an output of 750–1500 ml/min. Step increases in abdominal pressure resulted in rapid emptying presumably from the liver with a time constant of 0.61±0.1SE sec. followed by slower flow from non-hepatic viscera. The filling time constant was 0.57±0.09SE sec. Splanchnic emptying shifted up to 650 ml blood. With emptying, the increased hepatic vein flow increases the blood pressure at its entry into the inferior vena cava (IVC) and abolishes the pressure gradient producing flow between the femoral vein and the IVC inducing blood pooling in the legs. The findings are important for exercise because the larger the Vbs the greater the perfusion of locomotor muscles. During asystolic cardiac arrest we calculate that appropriate timing of abdominal compression could produce an output of 6 L/min. so that the abdominal circulatory pump might act as an auxiliary heart.
Expulsive maneuvers (EMs) caused by simultaneous contraction of diaphragm and abdominal muscles shift substantial quantities of blood from the splanchnic circulation to the extremities. This suggests that the diaphragm assisted by abdominal muscles might accomplish ventilation and circulation simultaneously by repeated EMs. We tested this hypothesis in normal subjects by measuring changes (Δ) in body volume (Vb) by whole body plethysmography simultaneously with changes in trunk volume (Vtr) by optoelectronic plethysmography, which measures the same parameters as whole body plethysmography plus the volume of blood shifts (Vbs) between trunk and extremities: Vbs = ΔVtr-ΔVb. We also measured abdominal pressure, pleural pressure, the arterial pressure wave, and cardiac output (Qc). EMs with abdominal pressure ~100 cmH(2)O for 1 s, followed by 2-s relaxations, repeated over 90 s, produced a "stroke volume" from the splanchnic bed of 0.35 ± 0.07 (SD) liter, an output of 6.84 ± 0.75 l/min compared with a resting Qc of 5.59 ± 1.14 l/min. Refilling during relaxation was complete, and the splanchnic bed did not progressively empty. Diastolic pressure increased by 25 mmHg during each EM. Between EMs, Qc increased to 7.09 ± 1.14 l/min due to increased stroke volume and heart rate. The circulatory function of the diaphragm assisted by simultaneous contractions of abdominal muscles with appropriate pressure and duration at 20 min(-1) can produce a circulatory output as great as resting Qc, as well as ventilation. These combined functions of the diaphragm have potential for cardiopulmonary resuscitation. The abdominal circulatory pump can act as an auxiliary heart.
Apart from its role as a flow generator for ventilation the diaphragm has a circulatory role. The cyclical abdominal pressure variations from its contractions cause swings in venous return from the splanchnic venous circulation. During exercise the action of the abdominal muscles may enhance this circulatory function of the diaphragm. Eleven healthy subjects (25 ± 7 year, 70 ± 11 kg, 1.78 ± 0.1 m, 3 F) performed plantar flexion exercise at ~4 METs. Changes in body volume (ΔVb) and trunk volume (ΔVtr) were measured simultaneously by double body plethysmography. Volume of blood shifts between trunk and extremities (Vbs) was determined non-invasively as ΔVtr-ΔVb. Three types of breathing were studied: spontaneous (SE), rib cage (RCE, voluntary emphasized inspiratory rib cage breathing), and abdominal (ABE, voluntary active abdominal expiration breathing). During SE and RCE blood was displaced from the extremities into the trunk (on average 0.16 ± 0.33 L and 0.48 ± 0.55 L, p < 0.05 SE vs. RCE), while during ABE it was displaced from the trunk to the extremities (0.22 ± 0.20 L p < 0.001, p < 0.05 RCE and SE vs. ABE respectively). At baseline, Vbs swings (maximum to minimum amplitude) were bimodal and averaged 0.13 ± 0.08 L. During exercise, Vbs swings consistently increased (0.42 ± 0.34 L, 0.40 ± 0.26 L, 0.46 ± 0.21 L, for SE, RCE and ABE respectively, all p < 0.01 vs. baseline). It follows that during leg exercise significant bi-directional blood shifting occurs between the trunk and the extremities. The dynamics and partitioning of these blood shifts strongly depend on the relative predominance of the action of the diaphragm, the rib cage and the abdominal muscles. Depending on the partitioning between respiratory muscles for the act of breathing, the distribution of blood between trunk and extremities can vary by up to 1 L. We conclude that during exercise the abdominal muscles and the diaphragm might play a role of an “auxiliary heart.”
This paper describes an integrated chest wearable apparatus for continuous blood pressure (BP) measurements exploiting the heart-rate (HR) and the pulse arrival time (PAT) in a modified Moens-Korteweg model. The device embeds a miniaturized gas pressure sensor to record the phonocardiogram (PCG) of the heart sounds, a LED-photodiode pair to detect the photoplethysmogram (PPG) of the blood flow wave, a µcontroller, a wireless communication module and the power supply. With the proposed device no active participation would be required from human subjects for BP measurements, since the HR and the PAT are continuously extracted from the PCG and PPG signals. Dedicated signal processing algorithms were developed and implemented off-line to extract both HR and PAT. A subject-specific calibration protocol of the BP model was designed and implemented. The calibration and validation of the apparatus were performed on a cohort of 20 healthy subjects. A GIMA ABPM pressure Holter was chosen as reference device, and 8 measurements points, evenly distributed over a 10-minute interval, were used for model calibration for each subject. The range of DBP and SBP measurements were 52-85 mmHg and 90-141 mmHg, respectively. The results from Bland-Altman analysis showed that the mean±1.96SD for the estimated diastolic, systolic, and mean BP with the proposed method against reference were 0.01±7.55, 1.47±3.76, 0.74±4.38 mmHg, respectively. The corresponding mean absolute differences (MAD) were 3.06, 1.83, and 1.80 mmHg. These results demonstrates that the acquisition apparatus is able to continuously estimate the BP with an accuracy comparable to traditional cuff-based devices. INDEX TERMS Wearable device, blood pressure measurement, pulse transit time (PTT), pulse arrival time (PAT), integrated system.
Electrical Impedance Tomography (EIT) is an image reconstruction technique applied in medicine for the electrical imaging of living tissues. In literature there is the evidence that a large resistivity variation related to the differences of the human tissues exists. As a result of this interest for the electrical characterization of the biological samples, recently the attention is also focused on the identification and characterization of the human tissue, by studying the homogeneity of its structure. An 8 electrodes needle-probe device has been developed with the intent of identifying the structural inhomogeneities under the surface layers. Ex-vivo impeditivity measurements, by placing the needle-probe in 5 different patterns of fat and lean porcine tissue, were performed, and impeditivity maps were obtained by EIDORS open source software for image reconstruction in electrical impedance. The values composing the maps have been analyzed, pointing out a good tissue discrimination, and the conformity with the real images. We conclude that this device is able to perform impeditivity maps matching to reality for position and orientation. In all the five patterns presented is possible to identify and replicate correctly the heterogeneous tissue under test. This new procedure can be helpful to the medical staff to completely characterize the biological sample, in different unclear situations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.