A pulsed quantum-cascade distributed-feedback laser, temperature tunable from -41 degrees C to +31.6 degrees C, and a resonant differential photoacoustic detector are used to measure trace-gas concentrations to as low as 66 parts per 10(9) by volume (ppbv) ammonia at a low laser power of 2 mW. Good agreement between the experimental spectrum and the simulated HITRAN spectrum of NH3 is found in the spectral range between 1046 and 1052 cm(-1). A detection limit of 30 ppbv ammonia at a signal-to-noise ratio of 1 was obtained with the quantum-cascade laser (QCL) photoacoustic (PA) setup. Concentration changes of approximately 50 ppbv were detectable with this compact and versatile QCL-based PA detection system. The performance of the PA detector, characterized by the product of the incident laser power and the minimum detectable absorption coefficient, was 4.7 x 10-9 W cm(-1).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.