A new dual-screen, dual-emulsion-film combination that allows a decrease in radiation dose of approximately 66% was compared with a widely used single-screen, single-emulsion-film system in contact and magnification mammography. Clustered microcalcifications randomly superimposed on a breast phantom were detected, and the location and number of individual calcifications were determined by four observers. The detectability of calcifications, determined with a receiver operating characteristic (ROC) analysis area, was 0.92 for magnification and 0.82 for contact mammography with the single-emulsion-film system, compared with 0.84 and 0.72, respectively, with the dual-emulsion-film system. More clusters were correctly located and more individual calcifications were counted with magnification than with contact mammography. The dual-emulsion-film system with the magnification technique performs as well as the single-emulsion-film system with the contact technique, while retaining a decrease in required dose of approximately 40%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.