The interferences of o-raffinose cross-linked hemoglobin (HemolinkTM) were examined and compared in two serum creatinine methods on the Hitachi 717 [Boehringer Mannheim (BMC) and Synermed] and in an enzymatic creatinine method on the Vitros 750 (Johnson & Johnson). Interference was considered significant when the change in creatinine concentration from the control exceeded the 95% confidence limits of each method. Significant interference was observed for the BMC/Hitachi 717 method with Hemolink ≥5 g/L. No interference was observed for the Synermed/Hitachi 717 assay with Hemolink up to 30 g/L in normal samples and up to 50 g/L in samples with increased creatinine. No interference was observed for the Vitros 750 assay with Hemolink up to 50 g/L at both normal and increased concentrations of creatinine. Although the BMC/Hitachi 717 method was considered unacceptable, the Synermed/Hitachi 717 and the Vitros 750 methods allow accurate quantification of serum creatinine in the presence of Hemolink.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.