We had presented an inductive type intraocular pressure sensor (L-sensor) in previous work. The distance between a micro coil and a ferrite on the membrane was modulated by pressure, and as a result the inductance and resonant frequency were changed. However, L-sensor has some problems to implant in eyes. First problem is low sensitivity. When L-sensor was implanted in rabbit's eyes, resonant frequency of L-sensor was very hard to detect. Second problem is biocompatibility. Size of L-sensor is 6×7×1.2 mm. When L-sensor was implanted in the eyes, it caused the inflammation. Therefore, this study suggests an inductive and capacitive type IOP sensor (LCsensor). The sensitivity of the LC-sensor 27.3 kHz/mmHg under 60 mmHg. It is much larger than 14 kHz/mmHg of the L-sensor. And the size of LC-sensor is 47% smaller than L-sensor. After 2 weeks from the implantation of LC-sensor into rabbit eyes, we measured the changes of resonant frequency of LC-sensor according to increased IOP by Balanced Salt Solution (BSS) injection. As a result, the sensitivity of LC-sensor in in vivo test is 25 kHz/mmHg. That is similar to the sensitivity of in vitro test. Received : Oct. 30, 2014, Revised : Nov. 28, 2014, Accepted : Nov. 28, 2014 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License(http://creativecommons.org/ licenses/bync/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.