Newborn screening for one or more lysosomal disorders has been implemented in several US states, Japan and Taiwan by multiplexed enzyme assays using either tandem mass spectrometry or digital microfluidics. Another multiplex assay making use of immunocapture technology has also been proposed. To investigate the potential variability in performance of these analytical approaches, we implemented three high-throughput screening assays for the simultaneous screening for four lysosomal disorders: Fabry disease, Gaucher disease, mucopolysaccharidosis type I, and Pompe disease. These assays were tested in a prospective comparative effectiveness study using nearly 100,000 residual newborn dried blood spot specimens. In addition, 2nd tier enzyme assays and confirmatory molecular genetic testing were employed. Post-analytical interpretive tools were created using the software Collaborative Laboratory Integrated Reports (CLIR) to determine its ability to improve the performance of each assay vs. the traditional result interpretation based on analyte-specific reference ranges and cutoffs. This study showed that all three platforms have high sensitivity, and the application of CLIR tools markedly improves the performance of each platform while reducing the need for 2nd tier testing by 66% to 95%. Moreover, the addition of disease-specific biochemical 2nd tier tests ensures the lowest false positive rates and the highest positive predictive values for any platform.
BACKGROUND
Friedreich ataxia (FRDA) is caused by reduced frataxin (FXN) concentrations. A clinical diagnosis is typically confirmed by DNA-based assays for GAA-repeat expansions or mutations in the FXN (frataxin) gene; however, these assays are not applicable to therapeutic monitoring and population screening. To facilitate the diagnosis and monitoring of FRDA patients, we developed an immunoassay for measuring FXN.
METHODS
Antibody pairs were used to capture FXN and an internal control protein, ceruloplasmin (CP), in 15 μL of whole blood (WB) or one 3-mm punch of a dried blood spot (DBS). Samples were assayed on a Luminex LX200 analyzer and validated according to standard criteria.
RESULTS
The mean recovery of FXN from WB and DBS samples was 99%. Intraassay and interassay imprecision (CV) values were 4.9%–13% and 9.8%–16%, respectively. The FXN limit of detection was 0.07 ng/mL, and the reportable range of concentrations was 2–200 ng/mL. Reference adult and pediatric FXN concentrations ranged from 15 to 82 ng/mL (median, 33 ng/mL) for DBS and WB. The FXN concentration range was 12–22 ng/mL (median, 15 ng/mL) for FRDA carriers and 1–26 ng/mL (median 5 ng/mL) for FRDA patients. Measurement of the FXN/CP ratio increased the ability to distinguish between patients, carriers, and the reference population.
CONCLUSIONS
This assay is applicable to the diagnosis and therapeutic monitoring of FRDA. This assay can measure FXN and the control protein CP in both WB and DBS specimens with minimal sample requirements, creating the potential for high-throughput population screening of FRDA.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.