Abstract. The automated laser-based hematology analyzer Sysmex XT-2000iV™ provides a 5-part differential count and specific cytograms that are of great interest for large veterinary laboratories. The aim of the study was to validate the Sysmex XT-2000iV compared to the laser-based hematology analyzer ADVIA® 2120 and manual differential in dogs, cats, and horses as well as the impact of anticoagulant (heparin, ethylenediamine tetra-acetic acid [EDTA], and citrate) and storage at 22°C and 4°C. Consecutive fresh K 3 -EDTA blood samples from 216 cats, 314 dogs, and 174 horses were included. The impact of anticoagulant and sample storage was assessed in specimens obtained from an additional 9 cats, 10 dogs, and 10 horses. Agreement between both analyzers was excellent to good except for monocytes and canine reticulocytes. Spearman rank correlation coefficients (r s ) between Sysmex XT-2000iV and manual differential were good to fair and ranged from 0.91 (cat lymphocytes) to 0.44 (cat monocytes). Hematocrit value (Hct), mean corpuscular hemoglobin (MCH), MCH concentration (MCHC; all: P < 0.001), and mean corpuscular volume (MCV; P < 0.01) were higher in canine citrated blood compared to heparin and EDTA. In cats, lymphocytes and monocytes were lower in heparinized blood compared to EDTA (P < 0.05), whereas in horses no significant effect was seen. Regarding storage time and temperature, white and red blood cell counts, hemoglobin, and MCH were stable. Hct, MCV, and MCHC were influenced by erythrocyte swelling. Differential count remained stable for 24 hr (22°C) and nearly 72 hr (4°C) except for monocytes. The overall performance of the Sysmex XT2000iV was excellent and compared favorably with that of the ADVIA 2120. A special strength was the excellent detection of feline eosinophils.
The automated laser-based hematology analyzer Sysmex XT-2000iV™ providing a complete blood cell count (CBC) and 5-part differential has been introduced in large veterinary laboratories. The aim of the current study was to determine precision, linearity, and accuracy of the Sysmex analyzer. Reference method for the accuracy study was the laser-based hematology analyzer ADVIA® 2120. For evaluation of accuracy, consecutive fresh blood samples from healthy and diseased cats (n = 216), dogs (n = 314), and horses (n = 174) were included. A low intra-assay coefficient of variation (CV) of approximately 1% was seen for the CBC except platelet count (PLT). An intra-assay CV ranging between 2% and 5.5% was evident for the differential count except for feline and equine monocytes (7.7%) and horse eosinophils (15.7%). Linearity was excellent for white blood cell count (WBC), hematocrit value, red blood cell count (RBC), and PLT. For all evaluated species, agreement was excellent for WBC and RBC, with Spearman rank correlation coefficients (r(s)) ranging from >0.99 to 0.98. Hematocrit value correlated excellently in cats and dogs, whereas for horses, a good correlation was evident. A good correlation between both analyzers was seen in feline and equine PLT (r(s) = 0.89 and 0.92, respectively), whereas correlation was excellent for dogs (r(s) = 0.93). Biases were close to 0 except for mean corpuscular hemoglobin concentration (4.11 to -7.25 mmol/l) and canine PLT (57 × 10(9)/l). Overall, the performance of the Sysmex analyzer was excellent and compared favorably with the ADVIA analyzer.
The PocH-100iV Diff is a suitable in-house instrument. A strength is its specific, but moderately sensitive, detection of feline eosinophils.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.