Digital tomosynthesis of the breast continues to be intensively studied as an adjunct or replacement of conventional mammography. Although many of the acquisition parameters found in tomosynthesis imaging are also found in conventional mammography and therefore most of the traditional values from mammography have been used in the former, two acquisition geometry parameters, the angular range covered during acquisition and the number of projections per projection set, are applicable only to tomosynthesis. Therefore, in the preclinical and clinical studies reported on tomosynthesis of the breast, a wide variety of values have been used for these two parameters. In this study, 63 different combinations of angular range and number of projections were evaluated using computer simulation methods to characterize how these two parameters affect reconstruction quality and to find which of these combinations maximize it. For this, a computer simulation of a digital tomosynthesis system that included empirically determined system response characteristics was developed and used to generate 9450 different breast tissue volume reconstructions. These reconstructions were analyzed objectively using metrics for in-plane lesion visibility and vertical resolution in the form of the contrast-to-noise ratio and artifact spread function (ASF). It was found that although maximizing the angular range covered does always increase the vertical resolution in tomosynthesis, increasing the number of projections in the acquisition set beyond a relatively low threshold does not further improve the vertical resolution. This threshold value for the minimal number of projections needed to minimize the ASF was found to vary proportionally with angular range. For example, for a 60 degrees angular range, the threshold number of projections was found to be 13. Given the clear inverse relationship between the number of projections and in-plane reconstruction quality under a limited total glandular dose condition, the optimum acquisition geometry in tomosynthesis imaging of the breast is that which maximizes the angular range while maintaining the number of projections close to the threshold values found. Finally, of the 63 acquisition geometries studied, the one that resulted in the highest quality reconstruction, considering both in-plane quality and vertical resolution, consisted of the acquisition of 13 projections over a 60 degrees angular range.
The purpose of this study was to investigate physical characteristics of two full field digital mammography (FFDM) systems (GE Senographe Essential and DS). Both are indirect conversion (x ray to light) alpha-Si flat panels coupled with a CsI(Tl) scintillator. The examined systems have the same pixel size (100 microm) but a different field of view: a conventional size 23 x 19.2 cm2 and a large field 24 X 30.7 cm2, specifically designed to image large breasts. In the GE Senographe Essential model relevant improvements in flat panel design were implemented and new deposition tools for metal, alpha-Si, and CsI(Tl) were introduced by GE. These changes in detector design are expected to be beneficial for advanced applications such as breast tomosynthesis. The presampling modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) were measured for a wide range of exposure (25-240 microGy) with a RQA-M2 technique (28 kVp with a Mo/Mo target/filter combination and 2 mm of additional aluminum filtration). At 1, 2, and at 4 lp/mm MTF is equal to 0.9, 0.76, and 0.46 for the conventional field detector and to 0.85, 0.59, and 0.24 for the large field detector. The latter detector exhibits an improved NNPS due to a lower electronic noise and a better DQE that reaches 60%. In addition a contrast-detail analysis was performed with CDMAM 3.4 phantom and CDCOM software: GE Senographe DS showed statistically significant poorer detection ability in comparison with the GE Senographe Essential. These results could have been expected, at least qualitatively, considering the relative DQE of the two systems.
Recently a new iterative reconstruction algorithm named Sinogram Affirmed Iterative Reconstruction (SAFIRE) has been released by Siemens. This algorithm works in the raw data domain with noise reduction as main purpose, providing five different strengths. In this study, the effect of SAFIRE on image quality has been investigated using selected phantoms and a comparison with standard filtered back projection (FBP) has been carried out. The following quantitative parameters have been evaluated: image noise, impact of different reconstruction kernels on noise reduction, noise power spectrum (NPS), contrast‐to‐noise ratio (CNR), spatial resolution, and linearity and accuracy of CT numbers. The influence of strengths on image quality parameters has also been examined. Results show that image noise reduction is independent of reconstruction kernel and strongly related to the strength of SAFIRE applied. The peak of NPS curve for SAFIRE reconstructions is shifted towards low frequencies; this effect is more marked at higher levels of strength. Contrast‐to‐noise ratio is always improved in SAFIRE reconstruction and increases with higher strength. At different dose levels SAFIRE preserves CT number accuracy, linearity, and spatial resolution, both in transversal and coronal planes. These results confirm that SAFIRE allows for image noise reduction with preserved image quality. First clinical data to validate this phantom analysis and confirm that commercially available iterative algorithms can play an effective role in dose containment.PACS number: 87.57.Q
The development of innovative nanosystems opens new perspectives for multidisciplinary applications at the frontier between materials science and nanomedicine. Here we present a novel hybrid nanosystem based on cytocompatible inorganic SiC/SiOx core/shell nanowires conjugated via click-chemistry procedures with an organic photosensitizer, a tetracarboxyphenyl porphyrin derivative. We show that this nanosystem is an efficient source of singlet oxygen for cell oxidative stress when irradiated with 6 MV X-Rays at low doses (0.4–2 Gy). The in-vitro clonogenic survival assay on lung adenocarcinoma cells shows that 12 days after irradiation at a dose of 2 Gy, the cell population is reduced by about 75% with respect to control cells. These results demonstrate that our approach is very efficient to enhance radiation therapy effects for cancer treatments.
Background Gallium-68 is a positron emitter for PET applications that can be produced without cyclotron by a germanium (Ge-68) chloride/gallium (Ga-68) chloride generator. Short half-life (67.71 min) of Ga-68, matching pharmacokinetic properties of small biomolecules, facilitates isotope utilization in compounding radiopharmaceuticals for PET imaging. The increasing cost of good manufacturing practice-compliant generators has strengthened the need for radionuclide efficient use by planning specific radiopharmaceutical sessions during the week, careful maintenance of the generator and achievement of high labeling yield and radiochemical purity (RCP) of the radiolabeled molecules. Methods The aim of this study was to evaluate the annual performance of five consecutive 68Ge/68Ga generators used for small-scale preparations of 68Ga-radiopharmaceuticals. To assess the long-term efficiency of isotope production we measured the weekly elution yield. To assess process efficiency we measured elution yield, labeling yield and RCP of four radiopharmaceutical preparations (68Ga-DOTATOC, 68Ga-PSMA-HBED-CC, 68Ga-PENTIXAFOR and 68Ga-DOTATATE). Results The annual mean elution yield of the generators was 74.7%, higher than that indicated by the manufacturer, and it never went below 65%. The Ge-68 level in the final products was under the detection limits in all the produced batches (mean value 0.0000048%). The RCP of radiopharmaceuticals determined by high-performance liquid chromatography was 98 ± 0.22%. The mean yield of radiolabelling was 64.68, 68.71, 57 and 63.68% for 68Ga-DOTATOC, 68Ga-PSMA-HBED-CC, 68GaPENTIXAFOR and 68Ga-DOTATATE. Conclusion The ability to prepare in the hospital radiopharmacy high-purity and pharmaceutically acceptable 68Ga-radiolabeled probes on a routine basis facilitates patient access to precision imaging for clinical and research aims.
In this study the regional cerebral blood flow (rCBF) pattern of three Mild Cognitive Impairment (MCI) subtypes was measured with SPECT in 60 patients (nineteen with an amnestic deficit, sixteen with disexecutive deficits, and twenty five with mild multidomain deficits) and compared with that of 15 healthy matched older adults. The amnestic MCI subgroup showed significant hypoperfusion in the left hippocampus, parahippocampal gyrus and fronto-parieto-temporal areas. The disexecutive subgroup had significant hypoperfusion of the left superior, medial frontal and cingulate cortex. The multidomain subgroup had similar perfusion deficits to the amnestic subgroup, with an additional deficit in the left posterior cingulate gyrus. This study found differential patterns of hypoperfusion in MCI subtypes. Since all patients who progressed to dementia converted to probable Alzheimer’s disease, the different rCBF patterns most likely reflect the neuropathological heterogeneity at onset and differences in disease stage.
Abstract. This study investigated the cognitive profile and the cerebral perfusion pattern in a highly educated 70 year old gentleman with posterior cortical atrophy (PCA). Visuo-perceptual abilities, spatial memory, spatial representation and navigation, visuospatial mental imagery, semantic and episodic-autobiographical memory were assessed. Regional cerebral blood flow (rCBF) was imaged with SPECT. Cognitive testing showed visual-perceptual impairment, apperceptive visual and landmark agnosia, topographical disorientation with way-finding deficits, impaired map learning and poor mental image generation. Semantic memory was normal, while episodic-autobiographical memory was impaired. Reduced rCBF was found mainly in the right hemisphere, in the precentral gyrus, posterior cingulate and middle temporal gyri, cuneus and precuneus, in the left superior temporal and lingual gyri and in the parahippocampus bilaterally. Hypoperfusion in occipito-parietal regions was associated with visuo-spatial deficits, whereas deficits in visuo-spatial mental imagery might reflect dysfunction related to hypoperfusion in the parahippocampus and precuneus, structures which are responsible for spatial and imagery processing. Dissociating performance between preserved semantic memory and poor episodic-autobiographical recall is consistent with a pattern of normal perfusion in frontal and anterior temporal regions but abnormal rCBF in the parahippocampi. The present findings indicate that PCA involves visuo-spatial imagery deficits and provide further validation to current neuro-cognitive models of spatial representation and topographical disorientation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.