A method of DNA immobilization on cellulose nitrate films has been developed. Modified films of uniform and stable surface have been used to devise two variants of solid-phase enzyme immunoassays of antibodies. The co-immobilization of enzyme label (cholinesterase) and the DNA molecules makes it possible to carry out the procedure of solid-phase enzyme immunoassay without any separation of components. Thus, it takes only 15 min to diagnose an autoimmune disease (Aleutian disease of minks) with the immunoenzyme amperometric sensor, with a lower detection limit for antibodies of 0.5 x 10(-10) M. For scaled diagnosing, solid-phase enzyme immunoassay on DNA-modified films with prior separation of components and spectrophotometric registration of peroxidase activity has been developed. The time for determination was 30 min, with a lower detection limit of 7.4 x 10(-12) M.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.