A forensic procedure for the quantitative confirmation of lysergic acid diethylamide (LSD) and the qualitative confirmation of its metabolite, N-demethyl-LSD, in blood, serum, plasma, and urine samples is presented. The Zymark RapidTrace™ was used to perform fully automated solid-phase extractions of all specimen types. After extract evaporation, confirmations were performed using liquid chromatography (LC) followed by positive electrospray ionization (ESI+) mass spectrometry/mass spectrometry (MS/MS) without derivatization. Quantitation of LSD was accomplished using LSD-d3 as an internal standard. The limit of quantitation (LOQ) for LSD was 0.05 ng/mL. The limit of detection (LOD) for both LSD and N-demethyl-LSD was 0.025 ng/mL. The recovery of LSD was greater than 95% at levels of 0.1 ng/mL and 2.0 ng/mL. For LSD at 1.0 ng/mL, the within-run and between-run (different day) relative standard deviation (RSD) was 2.2% and 4.4%, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.