This study compares the ability of 2 semiautomated methods with a fully automated method for QT measurement to minimize the sample size required to detect a moxifloxacin effect and exclude a placebo effect in a thorough QT/QTc study. The fully automated and 1 of 2 semiautomated methods used a global QT measurement in 12 leads, whereas the other semiautomated method used a tangent method on single lead raw complexes. Mean QTcF intervals were greater when measured on a global QT electrocardiogram than on raw complexes, but the mean magnitudes of DeltaQTcF were similar for all methods. The 3 methods detected a statistically significant increase in QTcF for moxifloxacin compared to placebo and were able to exclude a placebo effect on QTcF in all 62 participants. However, due to a smaller variability, the semiautomated methods allowed these detections with fewer than 20 participants, whereas the fully automated required at least 27 participants.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.