<b>Objective: </b>The use of remote
real-time continuous glucose monitoring (CGM) in the hospital has rapidly
emerged to preserve personal protective equipment (PPE) and reduce potential exposures
during COVID-19.
<p><b>Design and Methods:</b> We linked a hybrid CGM and point-of-care (POC)
glucose testing protocol to a computerized decision support system for
continuous insulin infusion (CII) and integrated a validation system for sensor
glucose values into the electronic health record. We report our
proof-of-concept experience in a COVID-19 ICU.<b></b></p>
<p><b>Results: </b>All nine patients
required mechanical ventilation and corticosteroids. Seventy six percent of
sensor values were within 20% of the reference POC glucose with an associated
average reduction in POC of 63%. Mean time in range (70-180 mg/dL) was 71.4 ± 13.9%.
Sensor accuracy was impacted by mechanical interferences in four patients.</p>
<p><b>Conclusions: </b>A hybrid
protocol integrating real-time CGM and POC is helpful for managing critically
ill patients with COVID-19 requiring insulin infusion. </p>
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.