Droplets in oil-water dispersions are measured using two different measurement techniques. Both 'focused beam reflectance measurements' (FBRM) and 'particle video microscopy' (PVM), are applied simultaneously in order to obtain direct comparison. A real time chord length distribution (CLD) is obtained from the FBRM. The picture series by the PVM are post-processed to obtain droplet size distributions (DSD). By measuring particle samples, it was shown that PVM is able to detect the correct size distribution.Using the PVM images, the uncertainty of the CLD was quantified. Two empirical approaches to convert the CLD to the correct DSD are documented. Downloaded by [University of Sherbrooke] at 12:09 12 April 2015 2
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.